Blood pressure control
Venous pressure:
Operating range
-700 to 750 mmHg
Alarm limits:
10 to 500 mmHg in treatment mode, -100
to 500 mmHg in priming mode
Accuracy:
-700 to -500 mmHg, ±10% -500 to 500
mmHg, ±5 mmHg or ±3%, whichever is
largest 500 to 750 mmHg, ±10%
Blood Pressure Monitor (BPM)
Systolic pressure range
1
60 - 250 mmHg
Low alarm limit, can be
preset (Default)
60 - 250 mmHg (100 mmHg)
High alarm limit, can be
preset (Default)
60 - 250 mmHg (150 mmHg)
Diastolic pressure range
1
40 - 200 mmHg
Low alarm limit, can be
preset (Default)
40 - 200 mmHg (60 mmHg)
High alarm limit, can be
preset (Default)
40 - 200 mmHg (100 mmHg)
Mean pressure range
1
45 - 235 mmHg
Low alarm limit, can be
preset (Default)
45 - 235 mmHg (75 mmHg)
High alarm limit, can be
preset (Default)
45 - 235 mmHg (115 mmHg)
Pulse rate range
40 - 200 bpm (±2 bpm or ±2% of reading)
Low alarm limit, can be
preset (Default)
40 - 200 bpm (40 bpm)
High alarm limit, can be
preset (Default)
40 - 200 bpm (150 bpm)
1
Meets ANSI/AAMI SP-10 (1992). Mean error ±5 mmHg. Standard
deviation 8 mmHg
H CEN9369 Revision.10.2005
Program version 12.xx
AK 95 S
Service
m anual - Technical data and spec i
fi
cations
3
:3
Summary of Contents for AK 95 S
Page 1: ...AK 95 AK 95 S Service Manual A better way to better care...
Page 5: ...This page is intentionally left blank...
Page 6: ...Installation Guide AK 95 AK 95 S January 2004...
Page 7: ......
Page 20: ......
Page 102: ...Maintenance Manual AK 95 AK 95 S October 2006 GAMBRO...
Page 103: ......
Page 125: ...5 22 MaintenanceManual AK95 AK95SHCEN12189Rev 10 2006 This page is intentionally left blank...
Page 126: ......
Page 152: ...Spare Parts List AK 95 AK 95 S October 2005...
Page 153: ......
Page 159: ...HCEN9094Rev 10 2005 7 3 Operator s panel 9 1 3 4 5 6 5 7 5 2 8 6 6 8...
Page 161: ...HCEN9094Rev 10 2005 7 4 Blood pump 10 2 8 3 5 9 4 7 1 6 11 12...
Page 163: ...HCEN9094Rev 10 2005 7 5 Blood pump Motor unit 8 7 6 9 2 10 3 4 5 1 11...
Page 165: ...HCEN9094Rev 10 2005 7 6 Heparin pump 1 2 3 4 5 6 8 7...
Page 169: ...HCEN9094Rev 10 2005 7 8 Magnetic Clamp unit 1 2 3 7 8 12 16 4 5 6 9 10 11 17 18 14 13 19 15...
Page 181: ...HCEN9094Rev 10 2005 7 13 Heater Flow guard 3 9 1 2 10 4 8 7 11 12 13 5 6 3...
Page 189: ...HCEN9094Rev 10 2005 Protective Protective 7 17 Boards 7 1 2 9 6 5 4 3 8...
Page 191: ...HCEN9094Rev 10 2005 1 7 18 Rear box 6 3 7 5 4 2...
Page 195: ...HCEN9094Rev 10 2005 T12AH 250AC BACH BACO HALT 7 20 AC DC 2 4 5 1 6 7 8 9 3...
Page 205: ...HCEN9094Rev 10 2005 7 25 MM I O board 3 2 9 4 6 7 8 5 1 10...
Page 215: ...HCEN9094Rev 10 2005 7 30 UFD Option 4 5 7 17 12 9 6 18 1 6 11 16 8 10 6 13 14 3 2 6 19 15...
Page 219: ...HCEN9094Rev 10 2005 7 32 Boards and Cablings EN1107 10 Step driver board...
Page 221: ...HCEN9094Rev 10 2005 7 32 Boards and Cablings EN1107 10 Step driver board...
Page 224: ......
Page 225: ...9 1 HCEN9160 Revision10 2005 Service Manual Symbols Chapter9 SymbolDescription...
Page 228: ...9 4 HCEN9160Revision10 2005 Service Manual Symbols Thispageisleftintentionallyblank...
Page 234: ...Replacements 10 6 AK 95 AK95SHCEN12078Rev 10 2005 This page is intentionally left blank...
Page 247: ...AK 95 AK 95 S Service Manual...