Technical Description
AK 95 / AK 95 S Service Manual
HCEN 9134 Rev. 10.2005
2:12
For example, if there is a blockage somewhere in
the tubes, the HPG generates an alarm and the
AK 95 S enters the bypass mode.
T
reg
after the heater is used together with the
temperature transducer on conductivity cell B,
to control the heater.
The bypass unit includes two valves:
The direct valve,
DIVA
, is normally open during
treatment (if the temperature, pressure, conduc-
tivity etc. are within the predefined limits).
The direct valve is closed and the bypass valve,
BYVA
, is opened when the dialysis fluid is
bypassed from the dialyzer. The dialysis fluid is
led past the dialyzer directly to the drain.
The dialysis fluid is forwarded to the dialyzer
via the output tube.
From the dialyzer the dialysis fluid (mixed with
waste products) is returned to the fluid unit via
the input tube. The input tube nipple is fitted
with a filter.
The
Dialysis fluid pressure transducer, PD,
after the dialyzer measures the dialysis fluid
pressure. This pressure transducer is the most
important pressure transducer in the system.
PD is used for calculation of the transmembrane
pressure,
TMP
.
TMP is calculated by the system and presented
on the operators panel. If the TMP is not within
the predefined limits, an alarm is issued.
The dialysis fluid is fed to a
deairating cham-
ber
, which is used to separate air from the fluid.
For example, a leakage may occur in the dialyzer
connection. It is important that the air bubbles
are separated from the dialysis fluid, otherwise
the UF measuring could be affected. When the
fluid level in the deairating chamber reaches the
minimum level, i.e. when there is too much air
in the deairating chamber, the evacuation valve,
EVVA
, is opened and the air is released. When
the fluid level reaches over the minimum level,
the evacuation valve is closed and so on. The
“degassed” fluid is taken out in the bottom of the
deairating chamber.
The taration valve,
TAVA
, is normally open
during treatment. During calibration of the UF
measuring unit, the taration valve is closed.
TAVA is also closed when the protective system
demands a minimum of UF.
The dialysis fluid is led through channel 2 of the
UF measuring cell to the
blood leak detector
.
The blood leak detector detects any blood in the
fluid. In the event of leak between the blood side
and dialysis fluid side of the dialyzer, an alarm
will be activated and the machine enters the
bypass mode.
Most of the dialysis fluid (90 to 95%) is taken out
at the top of the blood leak detector. The rest is
let out via a
drain restrictor
. The reason for
this is that the blood leak detector has to be
completely emptied during drain phase.
Finally, the dialysis fluid is led to the drain, via
the
suction pump
, the
outlet restrictor
and
the
drain-tube
.
Summary of Contents for AK 95 S
Page 1: ...AK 95 AK 95 S Service Manual A better way to better care...
Page 5: ...This page is intentionally left blank...
Page 6: ...Installation Guide AK 95 AK 95 S January 2004...
Page 7: ......
Page 20: ......
Page 102: ...Maintenance Manual AK 95 AK 95 S October 2006 GAMBRO...
Page 103: ......
Page 125: ...5 22 MaintenanceManual AK95 AK95SHCEN12189Rev 10 2006 This page is intentionally left blank...
Page 126: ......
Page 152: ...Spare Parts List AK 95 AK 95 S October 2005...
Page 153: ......
Page 159: ...HCEN9094Rev 10 2005 7 3 Operator s panel 9 1 3 4 5 6 5 7 5 2 8 6 6 8...
Page 161: ...HCEN9094Rev 10 2005 7 4 Blood pump 10 2 8 3 5 9 4 7 1 6 11 12...
Page 163: ...HCEN9094Rev 10 2005 7 5 Blood pump Motor unit 8 7 6 9 2 10 3 4 5 1 11...
Page 165: ...HCEN9094Rev 10 2005 7 6 Heparin pump 1 2 3 4 5 6 8 7...
Page 169: ...HCEN9094Rev 10 2005 7 8 Magnetic Clamp unit 1 2 3 7 8 12 16 4 5 6 9 10 11 17 18 14 13 19 15...
Page 181: ...HCEN9094Rev 10 2005 7 13 Heater Flow guard 3 9 1 2 10 4 8 7 11 12 13 5 6 3...
Page 189: ...HCEN9094Rev 10 2005 Protective Protective 7 17 Boards 7 1 2 9 6 5 4 3 8...
Page 191: ...HCEN9094Rev 10 2005 1 7 18 Rear box 6 3 7 5 4 2...
Page 195: ...HCEN9094Rev 10 2005 T12AH 250AC BACH BACO HALT 7 20 AC DC 2 4 5 1 6 7 8 9 3...
Page 205: ...HCEN9094Rev 10 2005 7 25 MM I O board 3 2 9 4 6 7 8 5 1 10...
Page 215: ...HCEN9094Rev 10 2005 7 30 UFD Option 4 5 7 17 12 9 6 18 1 6 11 16 8 10 6 13 14 3 2 6 19 15...
Page 219: ...HCEN9094Rev 10 2005 7 32 Boards and Cablings EN1107 10 Step driver board...
Page 221: ...HCEN9094Rev 10 2005 7 32 Boards and Cablings EN1107 10 Step driver board...
Page 224: ......
Page 225: ...9 1 HCEN9160 Revision10 2005 Service Manual Symbols Chapter9 SymbolDescription...
Page 228: ...9 4 HCEN9160Revision10 2005 Service Manual Symbols Thispageisleftintentionallyblank...
Page 234: ...Replacements 10 6 AK 95 AK95SHCEN12078Rev 10 2005 This page is intentionally left blank...
Page 247: ...AK 95 AK 95 S Service Manual...