AK 95 / AK 95 S Service Manual
HCEN 9134 Rev. 10.2005
Technical Description
2:25
Figure 14. Block diagram, power supply
Block Diagram Description
The mains voltage is fed to an AC/DC converter
which generates different DC supplies to the
monitor.
•Two +5 volt supplies, one for the control system
and the other for protective digital electronics.
•±12 volt is used for analog electronics.
•+24 volt is used for lamps, displays, blood-
pump, heparin pump, clamps, pumpmotor
(flow pump, suction pump), valves, fan and the
heating relay.
• The lower +12 V is a standby voltage for the
ON/OFF logic.
The mains voltage also supplies the voltage to
the heating element.
In a mains power failure situation, the AK 95 S
gives an intermittent buzzer alarm and performs
a recovery, when the power returns. All the
settings and accumulated values are stored in
the machine. The treatment will be continued
from where it was interrupted.
AK 95 S will perform a recovery when the
monitor is in treatment or disinfection.
In service and preset mode the monitor will start
with a Fch.
AC
DC
Battery
24 V
External fuses
Internal fuses
MAIN SWITCH
MAINS INPUT
MAINS to heater rod
12AH
12AH
+5VDC
+5VDC
±12VDC
+24VDC
+12VDC
AC
DC
Filter
Filter
Summary of Contents for AK 95 S
Page 1: ...AK 95 AK 95 S Service Manual A better way to better care...
Page 5: ...This page is intentionally left blank...
Page 6: ...Installation Guide AK 95 AK 95 S January 2004...
Page 7: ......
Page 20: ......
Page 102: ...Maintenance Manual AK 95 AK 95 S October 2006 GAMBRO...
Page 103: ......
Page 125: ...5 22 MaintenanceManual AK95 AK95SHCEN12189Rev 10 2006 This page is intentionally left blank...
Page 126: ......
Page 152: ...Spare Parts List AK 95 AK 95 S October 2005...
Page 153: ......
Page 159: ...HCEN9094Rev 10 2005 7 3 Operator s panel 9 1 3 4 5 6 5 7 5 2 8 6 6 8...
Page 161: ...HCEN9094Rev 10 2005 7 4 Blood pump 10 2 8 3 5 9 4 7 1 6 11 12...
Page 163: ...HCEN9094Rev 10 2005 7 5 Blood pump Motor unit 8 7 6 9 2 10 3 4 5 1 11...
Page 165: ...HCEN9094Rev 10 2005 7 6 Heparin pump 1 2 3 4 5 6 8 7...
Page 169: ...HCEN9094Rev 10 2005 7 8 Magnetic Clamp unit 1 2 3 7 8 12 16 4 5 6 9 10 11 17 18 14 13 19 15...
Page 181: ...HCEN9094Rev 10 2005 7 13 Heater Flow guard 3 9 1 2 10 4 8 7 11 12 13 5 6 3...
Page 189: ...HCEN9094Rev 10 2005 Protective Protective 7 17 Boards 7 1 2 9 6 5 4 3 8...
Page 191: ...HCEN9094Rev 10 2005 1 7 18 Rear box 6 3 7 5 4 2...
Page 195: ...HCEN9094Rev 10 2005 T12AH 250AC BACH BACO HALT 7 20 AC DC 2 4 5 1 6 7 8 9 3...
Page 205: ...HCEN9094Rev 10 2005 7 25 MM I O board 3 2 9 4 6 7 8 5 1 10...
Page 215: ...HCEN9094Rev 10 2005 7 30 UFD Option 4 5 7 17 12 9 6 18 1 6 11 16 8 10 6 13 14 3 2 6 19 15...
Page 219: ...HCEN9094Rev 10 2005 7 32 Boards and Cablings EN1107 10 Step driver board...
Page 221: ...HCEN9094Rev 10 2005 7 32 Boards and Cablings EN1107 10 Step driver board...
Page 224: ......
Page 225: ...9 1 HCEN9160 Revision10 2005 Service Manual Symbols Chapter9 SymbolDescription...
Page 228: ...9 4 HCEN9160Revision10 2005 Service Manual Symbols Thispageisleftintentionallyblank...
Page 234: ...Replacements 10 6 AK 95 AK95SHCEN12078Rev 10 2005 This page is intentionally left blank...
Page 247: ...AK 95 AK 95 S Service Manual...