11-7
No Check
Items
Check
Procedure
Criteria
9. NIBP
01
NIBP Test
Connect the 500 ml tank, and perform the NIBP
test under the test menu.
All the test results should be "OK".
02 BP
Measurement
Error
Set the simulator to 120 mmHg for SYS, 80
mmHg for DIA, 90 mmHg for MAP, and perform
the NIBP measurement.
The displayed BP value should be within 120
mmHg ± 10 mmHg for SYS, 80 mmHg ±10
mmHg for DIA, 90 mmHg ±10 mmHg for
MAP.
03 Pulse
Rate
Measurement Error
Set the pulse rate to 60 bpm on the simulator,
and perform the NIBP measurement.
The displayed pulse rate should be within 60
bpm ± 3 bpm.
No Check
Items
Check
Procedure
Criteria
10. Temperature
01 Temperature
display
error
Input the temperature signal of 37°C from the
simulator, and check the displayed temperature
value on the monitor.
The displayed temperature value should be
within 37°C ± 0.2°C.
No Check
Items
Check
Procedure
Criteria
11. Cardiac Output (Blood Temperature, Injectate Temperature)
01 Blood
Temperature
Measurement
Input the blood temperature signal of 37°C from
the simulator, and check the displayed blood
temperature value on the monitor.
The displayed blood temperature value
should be within 37°C ± 0.3°C.
02 Injectate
Temperature
Measurement
Input the injectate temperature signal of 15°C
from the simulator, and check the displayed
injectate temperature value on the monitor.
The displayed injectate temperature value
should be within 15°C ± 0.5°C.
No Check
Items
Check
Procedure
Criteria
12. CO
2
Concentration (Optional)
01
CO
2
Concentration
Calibration
(HCP-810/820)
Perform calibration according to the procedure
explained in "Chapter 9 Maintenance
Check"(Maintenance Manual).
The calibration should complete without
error.
02
CO
2
Concentration
Measurement
Perform measurement with 5% calibration gas,
and check the displayed CO
2
concentration
value on the monitor.
The displayed CO
2
concentration value
should be within 38 mmHg ± 2 mmHg.
No Check
Items
Check
Procedure
Criteria
13. BIS (Optional)
01
Sensor Detection
Connect the signal generator and BIS sensor
simulator, and check that EEG is displayed.
EEG should be displayed on the monitor.
No Check
Items
Check
Procedure
Criteria
14. Analog Output
01
ECG Output Accuracy Input 10 Hz sine-wave signal of 1mV (2mVp-p)
amplitude from the simulator, connect the
oscilloscope to the output, and check the
displayed waveform.
The displayed waveform amplitude should
be within 1 V ± 0.1 V (2 Vp-p ± 0.2 Vp-p).
02 BP
Output
Accuracy Input BP signal of 0 mmHg, 100 mmHg, 250
mmHg from the simulator, connect the
oscilloscope to the output, and check the
displayed waveform.
The displayed output voltage should be
within 0 V ± 0.1 V at 0 mmHg,
The displayed output voltage should be
within 1 V ± 0.1 V at 100 mmHg,
Summary of Contents for DS-8400
Page 4: ......
Page 36: ...xxxii ...
Page 46: ...1 6 ...
Page 74: ...3 4 ...
Page 117: ...5 1 Wiring Diagram Chapter 5 Main Unit DSC 8400 5 2 ...
Page 120: ...5 4 ...
Page 128: ...6 8 ...
Page 130: ...7 2 DSC 8410 CPU Unit BLK ...
Page 132: ...7 4 DSC 8410 Power Supply Unit BLK ...
Page 135: ...7 7 DSC 8410 Assembly 1 Front Case Part ...
Page 137: ...7 9 DSC 8410 Assembly 2 CPU Unit Part ...
Page 139: ...7 11 DSC 8410 Assembly 3 Joint Part Expansion Module Part ...
Page 141: ...7 13 DSC 8410 Assembly 4 COM AUX Part Recorder Part ...
Page 143: ...7 15 DSC 8410 Assembly 5 Power Board Part Power Supply Unit Part ...
Page 145: ...7 17 DSC 8410 Assembly 6 Rear Case Sheet Metal Part Item number ...
Page 147: ...7 19 DSC 8410 Assembly 7 Rear Case Part Item number Connector position ...
Page 149: ...7 21 DSC 8410 Assembly 8 Front Rear Stand Part ...
Page 152: ...7 24 DSC 8410 Assembly 10 Super Unit Cover Assembly ...
Page 154: ...7 26 DSC 8410 Assembly 11 Super Unit Part Blank Cover Recorder Part Blank ...
Page 156: ...7 28 DSC 8410 Assembly 12 Labeling ...
Page 168: ...8 6 ...
Page 222: ...9 54 ...
Page 258: ...10 36 ...
Page 272: ...11 14 ...