xvii
pacemaker pulse amplitude to decrease, and disables the pacemaker pulse detection.
If signals similar to a pacemaker pulse are present, such as electric blanket noise or excessive AC frequency
noise, these may be erroneously detected and displayed as a pacemaker pulse.
When a spontaneous QRS and pacemaker pulse overlap (ex. fusion beat, etc.), QRS detection cannot be
performed properly. In this case, the heart rate is degraded.
If a pacemaker pulse is continuously detected due to AC frequency interference, QRS detection will be
suspended and the heart rate will be reduced. Arrhythmia will not be detected either.
Precautions about the ST Measurement
The ST algorithm has been tested for accuracy of the ST segment data. The significance of the ST
segment changes need to be determined by a clinician.
For the lead which the electrode is detached, the reference waveform setup cannot be performed. Check if the
electrode is appropriately attached, and perform the setup again.
Precautions about the 12-Lead Analysis
Interpretation and Minnesota codes given by this equipment do not instruct the physician as to the kind and
degree of cardiac disease. Accordingly, four value judgments are given for ECG waveform and although
"abnormal" indicates a large possibility of organic cardiac disease, there are cases where no cardiac disease
exists despite an abnormal ECG (that is, an abnormal ECG may be caused by something other than heart).
On the other hand, care should be taken in the event any preclinical coronary arteriosclerosis could be present
despite a normal ECG interpretation.
Therefore, for a proper diagnosis, the ECG should be integrated with other interpretations.
ECG Recording by the Mason-Likar System
The 12-lead ECG recorded with the torso placement of the limb leads (Mason-Likar 12-lead system) may differ
from the standard 12-lead ECG. Moreover, waveforms may differ somewhat also in a supine position and a
standing position (sitting position).
Fukuda Denshi recommends to carry out the recording of the ECG by taking into consideration the waveform
differences according to electrode positions or postures.
For the model installed with ECG analysis program
The ECG analysis program is intended to analyze the standard 12-lead ECG waveforms. Therefore, the
analyzed result for waveforms recorded with the torso placement of the limb leads (Mason-Likar 12-lead
system) may differ from that of the standard 12-lead ECG waveforms.
Select "Used" for the pacemaker setting on the patient admit/discharge menu if a patient has a pacemaker.
The threshold values for classification of 12-lead ECG interpretation and Minnesota code are set by age and
sex as follows:
1.
Male and Female of ages 19 years old and above
2.
Male of age 12 through 18 years old
3.
Female of age 12 through 18 years old
4.
Male and Female of ages 3 through 11 years old
5.
Male and Female of ages below 2 years old
If no patient information (i.e. Default : "Class." [Adult], "Sex": undetermined, and "Age" [0]) has been entered,
the system algorithm will handle the patient as a "35 years old male".
Before the analysis, make sure the patient classification ([Adult] / [Child]) is properly selected. If [Neonate] is
selected, the 12-lead ECG analysis will not function.
Enter the age of patient if known. If no age information (i.e. Default: [0]) has been entered, the
system algorithm will handle the patient as "35 years old".
Enter the sex of patient if known. If no sex information (i.e. Default: undetermined) has been entered, the
system algorithm will handle the patient as "Male".
If the patient classification is set as [Child] and no age (i.e. Default: [0] ) has been entered, the system algorithm
will handle the patient as "less than 2 years old""
Precautions about the SpO
2
Monitoring
Use only the sensor/relay cable specified by Fukuda Denshi. Otherwise, it may cause measurement error.If
the sensor is damaged, stop using it.
If the nail is rough, dirty, or manicured, accurate measurement will not be possible. Change the finger or clean
the nail before attaching the sensor.
CAUTION
CAUTION
CAUTION
Summary of Contents for DS-8400
Page 4: ......
Page 36: ...xxxii ...
Page 46: ...1 6 ...
Page 74: ...3 4 ...
Page 117: ...5 1 Wiring Diagram Chapter 5 Main Unit DSC 8400 5 2 ...
Page 120: ...5 4 ...
Page 128: ...6 8 ...
Page 130: ...7 2 DSC 8410 CPU Unit BLK ...
Page 132: ...7 4 DSC 8410 Power Supply Unit BLK ...
Page 135: ...7 7 DSC 8410 Assembly 1 Front Case Part ...
Page 137: ...7 9 DSC 8410 Assembly 2 CPU Unit Part ...
Page 139: ...7 11 DSC 8410 Assembly 3 Joint Part Expansion Module Part ...
Page 141: ...7 13 DSC 8410 Assembly 4 COM AUX Part Recorder Part ...
Page 143: ...7 15 DSC 8410 Assembly 5 Power Board Part Power Supply Unit Part ...
Page 145: ...7 17 DSC 8410 Assembly 6 Rear Case Sheet Metal Part Item number ...
Page 147: ...7 19 DSC 8410 Assembly 7 Rear Case Part Item number Connector position ...
Page 149: ...7 21 DSC 8410 Assembly 8 Front Rear Stand Part ...
Page 152: ...7 24 DSC 8410 Assembly 10 Super Unit Cover Assembly ...
Page 154: ...7 26 DSC 8410 Assembly 11 Super Unit Part Blank Cover Recorder Part Blank ...
Page 156: ...7 28 DSC 8410 Assembly 12 Labeling ...
Page 168: ...8 6 ...
Page 222: ...9 54 ...
Page 258: ...10 36 ...
Page 272: ...11 14 ...