Precautions when Using with Other Equipment
Pacemaker
Minute ventilation rate-adaptive implantable pacemakers can occasionally interact with certain cardiac
monitoring and diagnostic equipment, causing the pacemakers to pace at their maximum programmed rate. The
cardiac monitoring and diagnostic equipment may possibly send wrong information. If such event occurs,
please disconnect the cardiac monitoring and diagnostic equipment, or follow the procedures described in the
operation manual of the pacemaker. For more details, contact FUKUDA DENSHI personnel, your institution's
professionals, or your pacemaker distributors.
Rate meters may continue to count the pacemaker rate during occurrences of cardiac arrest or some
arrhythmias. Do not rely entirely upon rate meter alarms. Keep pacemaker patients under close surveillance.
Reference
"Minute Ventilation Rate-Adaptive Pacemakers"
FDA alerts health professionals that minute ventilation rate-adaptive implantable pacemakers can
occasionally interact with certain cardiac monitoring and diagnostic equipment, causing pacemakers to pace
at their maximum programmed rate.
[Based on a safety bulletin issued by FDA Center for Devices and Radiological Health on October 14, 1998]
Non-Explosion Proof
Never operate the equipment in the presence of flammable anesthetics, high concentration of oxygen, or
inside hyperbaric chamber. Also, do not operate the equipment in an environment in which there is a risk of
explosion. Explosion or fire may result.
Defibrillator
When defibrillating, keep away from the electrodes or medicament applied to the patient chest. If this is not
possible, remove the electrodes or medicament before defibrillating.
If the defibrillator paddles are directly in contact with the electrodes or medicament, an electrical shock may
result by the discharged energy.
When defibrillating, make sure that the electrodes, sensor cables, or relay cables are firmly connected to the
device.
Contacting the metal part of the disconnected cable may result in electrical shock from the discharged energy.
When defibrillating, do not touch the patient and the metal part of the device or cables.
Electric shock may result from the discharged energy.
This equipment will return to standard operating mode within 10 seconds after defibrillating. However, when
in diagnosis mode, it may require 10 seconds or more after defibrillation to display the normal ECG waveform
as the time constant setting is large.
The stored data will not be affected. The measurement accuracy will temporarily decrease during defibrillation,
but it will not compromise the safety of patient and the equipment.
The QRS synchronized signal is not intended to be used as synchronized signal for defibrillator.
WARNING
DANGER
WARNING
Summary of Contents for DS-8400
Page 4: ......
Page 36: ...xxxii ...
Page 46: ...1 6 ...
Page 74: ...3 4 ...
Page 117: ...5 1 Wiring Diagram Chapter 5 Main Unit DSC 8400 5 2 ...
Page 120: ...5 4 ...
Page 128: ...6 8 ...
Page 130: ...7 2 DSC 8410 CPU Unit BLK ...
Page 132: ...7 4 DSC 8410 Power Supply Unit BLK ...
Page 135: ...7 7 DSC 8410 Assembly 1 Front Case Part ...
Page 137: ...7 9 DSC 8410 Assembly 2 CPU Unit Part ...
Page 139: ...7 11 DSC 8410 Assembly 3 Joint Part Expansion Module Part ...
Page 141: ...7 13 DSC 8410 Assembly 4 COM AUX Part Recorder Part ...
Page 143: ...7 15 DSC 8410 Assembly 5 Power Board Part Power Supply Unit Part ...
Page 145: ...7 17 DSC 8410 Assembly 6 Rear Case Sheet Metal Part Item number ...
Page 147: ...7 19 DSC 8410 Assembly 7 Rear Case Part Item number Connector position ...
Page 149: ...7 21 DSC 8410 Assembly 8 Front Rear Stand Part ...
Page 152: ...7 24 DSC 8410 Assembly 10 Super Unit Cover Assembly ...
Page 154: ...7 26 DSC 8410 Assembly 11 Super Unit Part Blank Cover Recorder Part Blank ...
Page 156: ...7 28 DSC 8410 Assembly 12 Labeling ...
Page 168: ...8 6 ...
Page 222: ...9 54 ...
Page 258: ...10 36 ...
Page 272: ...11 14 ...