9-18
C02-02 The air hose was disconnected from the NIBP connector during the measurement.
Cause
The air hose was disconnected from the NIBP connector during the measurement.
Solution
Connect the air hose to the NIBP connector, and then measure again.
C03-xx The exhaust ventilation has ceased, or the target deflation speed was not achieved.
Cause 1
During measurement, an artifact such as body motion may have interfered.
Solution
Keep the patient still as much as possible, and measure while the patient is not moving. When performing the
measurement during surgery, avoid artifact caused by the surgery.
Cause 2
During the measurement, air hose was bent or occluded by the compression.
Solution
Make sure that the air hose is not bent or compressed before the measurement.
If the error persists and C03-xx error is frequently displayed, contact Fukuda Denshi and notify the error code.
C04-xx The cuff inflation was insufficient for the patient's blood pressure.
Cause
The blood pressure has significantly increased from the previous measurement.
Solution
Check the cuff application and size and perform the manual measurement.
C06-xx The pulse signal detected during the measurement was unstable.
Cause 1
During the measurement, the patient has trembled or moved.
Solution
Keep the patient still as much as possible, and measure while the patient is not trembling or moving.
Cause 2
Arrhythmia has frequently occurred during the measurement.
Solution
If arrhythmia occurs many times, correct measurement cannot be performed. Measure when arrhythmia is not
frequently occurring.
C07-00 The measurement time has exceeded the allowable time.
Cause
Measurement is automatically repeated due to body motion or insufficient inflation.
Solution
Check the cuff application and size, and measure while keeping the patient still as much as possible.
Summary of Contents for DS-8400
Page 4: ......
Page 36: ...xxxii ...
Page 46: ...1 6 ...
Page 74: ...3 4 ...
Page 117: ...5 1 Wiring Diagram Chapter 5 Main Unit DSC 8400 5 2 ...
Page 120: ...5 4 ...
Page 128: ...6 8 ...
Page 130: ...7 2 DSC 8410 CPU Unit BLK ...
Page 132: ...7 4 DSC 8410 Power Supply Unit BLK ...
Page 135: ...7 7 DSC 8410 Assembly 1 Front Case Part ...
Page 137: ...7 9 DSC 8410 Assembly 2 CPU Unit Part ...
Page 139: ...7 11 DSC 8410 Assembly 3 Joint Part Expansion Module Part ...
Page 141: ...7 13 DSC 8410 Assembly 4 COM AUX Part Recorder Part ...
Page 143: ...7 15 DSC 8410 Assembly 5 Power Board Part Power Supply Unit Part ...
Page 145: ...7 17 DSC 8410 Assembly 6 Rear Case Sheet Metal Part Item number ...
Page 147: ...7 19 DSC 8410 Assembly 7 Rear Case Part Item number Connector position ...
Page 149: ...7 21 DSC 8410 Assembly 8 Front Rear Stand Part ...
Page 152: ...7 24 DSC 8410 Assembly 10 Super Unit Cover Assembly ...
Page 154: ...7 26 DSC 8410 Assembly 11 Super Unit Part Blank Cover Recorder Part Blank ...
Page 156: ...7 28 DSC 8410 Assembly 12 Labeling ...
Page 168: ...8 6 ...
Page 222: ...9 54 ...
Page 258: ...10 36 ...
Page 272: ...11 14 ...