9-3
If the lead cable or relay cable is defective (wire break, etc.), replace it.
If any noise source is near the patient, move it away from the patient as far as possible.
Cause 2
EMG is interfering.
Solution
Change the electrode site to a location where the myoelectricity will be less likely to interfere.
Select ESIS for the filter mode.
Selecting ESIS for the filter mode will decrease the QRS amplitude and may result in
not counting the heart rate.
Cause 3
Even though the "3-lead Override" (ECG Parameter Setup) is set to [ON], electrodes other than L (LA), R (RA), F
(LL) are connected.
Solution
Set the "3-lead Override" to [OFF].Or, detach the electrodes other than L (LA), R (RA), F (LL).
The ECG waveform is in the baseline position.
The lead-off condition may have occurred by the following causes.
Cause 1
Electrode is detached.
Solution
Place the electrodes again.If the electrode contact is poor, replace the electrode.
Cause 2
The lead cable is disconnected from the electrode terminal.
Solution
Securely connect the lead cable.
If the error persists, wire break of the lead cable or relay cable can be considered.
<Check Electrodes Attachment> is displayed.
Cause 1
The electrode contact with the skin is poor. There is substantial contact resistance between the electrodes.
Solution
Replace all the electrodes.Make sure to use the same type of electrodes .
Cause 2
Even though the "3-lead Override" (ECG Parameter Setup) is set to [ON], electrodes other than L (LA), R (RA), F
(LL) are connected.
Solution
Set the "3-lead Override" to [OFF]. Or, detach the electrodes other than L (LA), R (RA), F (LL).
REFERENCE
CAUTION
Summary of Contents for DS-8400
Page 4: ......
Page 36: ...xxxii ...
Page 46: ...1 6 ...
Page 74: ...3 4 ...
Page 117: ...5 1 Wiring Diagram Chapter 5 Main Unit DSC 8400 5 2 ...
Page 120: ...5 4 ...
Page 128: ...6 8 ...
Page 130: ...7 2 DSC 8410 CPU Unit BLK ...
Page 132: ...7 4 DSC 8410 Power Supply Unit BLK ...
Page 135: ...7 7 DSC 8410 Assembly 1 Front Case Part ...
Page 137: ...7 9 DSC 8410 Assembly 2 CPU Unit Part ...
Page 139: ...7 11 DSC 8410 Assembly 3 Joint Part Expansion Module Part ...
Page 141: ...7 13 DSC 8410 Assembly 4 COM AUX Part Recorder Part ...
Page 143: ...7 15 DSC 8410 Assembly 5 Power Board Part Power Supply Unit Part ...
Page 145: ...7 17 DSC 8410 Assembly 6 Rear Case Sheet Metal Part Item number ...
Page 147: ...7 19 DSC 8410 Assembly 7 Rear Case Part Item number Connector position ...
Page 149: ...7 21 DSC 8410 Assembly 8 Front Rear Stand Part ...
Page 152: ...7 24 DSC 8410 Assembly 10 Super Unit Cover Assembly ...
Page 154: ...7 26 DSC 8410 Assembly 11 Super Unit Part Blank Cover Recorder Part Blank ...
Page 156: ...7 28 DSC 8410 Assembly 12 Labeling ...
Page 168: ...8 6 ...
Page 222: ...9 54 ...
Page 258: ...10 36 ...
Page 272: ...11 14 ...