9-14
<SpO
2
Pulse Search> is displayed.
Cause 1
The amplitude of the pulse waveform is too low, or the sensor is not positioned correctly.
Solution
Check that the light emitting and receiving parts of the sensor LED are aligned.
Cause 2
The sensor has not been attached long enough to obtain stable measurement.
Solution
After the sensor attachment, wait and see for about one minute until the waveform stabilizes.
<SpO
2
Noise Interference> is displayed.
Cause
External signal or energy is interfering with the measurement.
Solution
Remove the external interference or apply ambient shielding.
<SpO
2
Check Sensor>, <SpO
2
Replace Cable>, or <SpO
2
Check Cable> is displayed.
Cause 1
Unrecognizable sensor is connected.
A wrong patient cable is used.
When attached to the patient, the sensor was exposed to high-intensity light which lead to false recognition.
Solution
Reattach the SpO
2
sensor and patient cable.
Replace with a Fukuda Denshi specified patient cable and sensor.
(
"Pulse Oximetry Measurement (Masimo)" P13-4)
Cause 2
The cable is used beyond its expected life.
Solution
Replace the patient cable.
The X-Cal function automatically monitors the Active Monitoring Time (actual time of SpO
2
monitoring) for each sensor and cable.
Even if the cable is used beyond its expected life, the measurement will not cease unless
the power is turned OFF or the cable is reconnected.
When a measurement with a cable that has reached its end of life is suspended for certain
amount of time, and resumed with the same cable, a message to replace the cable will be
displayed.
Depending on the equipment, some cable may not be recognized.
NOTE
Summary of Contents for DS-8400
Page 4: ......
Page 36: ...xxxii ...
Page 46: ...1 6 ...
Page 74: ...3 4 ...
Page 117: ...5 1 Wiring Diagram Chapter 5 Main Unit DSC 8400 5 2 ...
Page 120: ...5 4 ...
Page 128: ...6 8 ...
Page 130: ...7 2 DSC 8410 CPU Unit BLK ...
Page 132: ...7 4 DSC 8410 Power Supply Unit BLK ...
Page 135: ...7 7 DSC 8410 Assembly 1 Front Case Part ...
Page 137: ...7 9 DSC 8410 Assembly 2 CPU Unit Part ...
Page 139: ...7 11 DSC 8410 Assembly 3 Joint Part Expansion Module Part ...
Page 141: ...7 13 DSC 8410 Assembly 4 COM AUX Part Recorder Part ...
Page 143: ...7 15 DSC 8410 Assembly 5 Power Board Part Power Supply Unit Part ...
Page 145: ...7 17 DSC 8410 Assembly 6 Rear Case Sheet Metal Part Item number ...
Page 147: ...7 19 DSC 8410 Assembly 7 Rear Case Part Item number Connector position ...
Page 149: ...7 21 DSC 8410 Assembly 8 Front Rear Stand Part ...
Page 152: ...7 24 DSC 8410 Assembly 10 Super Unit Cover Assembly ...
Page 154: ...7 26 DSC 8410 Assembly 11 Super Unit Part Blank Cover Recorder Part Blank ...
Page 156: ...7 28 DSC 8410 Assembly 12 Labeling ...
Page 168: ...8 6 ...
Page 222: ...9 54 ...
Page 258: ...10 36 ...
Page 272: ...11 14 ...