H3-28
006-252-02
10.22.99 FM2569
FM-DP L
Service Manual
H3-28
MAINTENANCE 3. TROUBLESHOOTING
Error code
Error name
Significance
Probable cause and remedy
Detailed
information format
0238
Sensor state 3 error
Attempted to move the sub scanning uni
grip from the home position to the grip
state during initialization in order to
detect/discharge the film(s) remaining in
the sub scanning unit, but SD2/SD3 did
not become CLOSE.
The following motor/sensor/part may be faulty. Correct
according to the troubleshooting procedure for this error code
in "3.1.4 Troubleshooting Flow".
SD2/SD3/MD2
2-1-C
2-2-A
0239
Sensor state 1 error
Attempted to move the sub scanning unit
grip from the grip state to the home
position during initialization in order to
detect/discharge the film(s) remaining in
the sub scanning unit, but SD2/SD3 did
not become OPEN.
The following motor/sensor/part may be faulty. Correct
according to the troubleshooting procedure for this error code
in "3.1.4 Troubleshooting Flow".
SD2/SD3/MD2
2-1-C
2-2-A
0240
SE1HP return error
Attempted to move the transport unit
grip to the home position during
initialization in order to detect/discharge
the film(s) remaining in the sub scanning
unit, but SE1 did not become CLOSE.
The following motor/sensor/part may be faulty. Correct
according to the troubleshooting procedure for this error code
in "3.1.4 Troubleshooting Flow".
SE1/ME1
2-1-C
2-2-A
0241
Sensor state 3 error
Attempted to move the sub scanning uni
grip from the home position to the grip
state during initialization in order to
detect/discharge the film(s) remaining in
the before-sub-scanning conveyor, but
SD2/SD3 did not become CLOSE.
The following motor/sensor/part may be faulty. Correct
according to the troubleshooting procedure for this error code
in "3.1.4 Troubleshooting Flow".
SD2/SD3/MD2
2-1-C
2-2-A
0242
Sensor state 1 error
Attempted to move the sub scanning unit
grip from the grip state to the home
position during initialization in order to
detect/discharge the film(s) remaining in
the before-sub-scanning conveyor, but
SD2/SD3 did not become OPEN.
The following motor/sensor/part may be faulty. Correct
according to the troubleshooting procedure for this error code
in "3.1.4 Troubleshooting Flow".
SD2/SD3/MD2
2-1-C
2-2-A
0244
SE1HP return error
Attempted to move the transport unit
grip to the home position during
initialization in order to detect/discharge
the film(s) remaining in the before-sub-
scanning conveyor, but SE1 did not
become CLOSE.
The following motor/sensor/part may be faulty. Correct
according to the troubleshooting procedure for this error code
in "3.1.4 Troubleshooting Flow".
SE1/ME1
2-1-C
2-2-A
0249
Tray 1
film surface detection
error
Attempted to move the suction cup unit
from the upper dead-point to the surface
detection position during removal at tray
1, but SA3 did not become CLOSE.
The following motor/sensor/part may be faulty. Correct
according to the troubleshooting procedure for this error code
in "3.1.4 Troubleshooting Flow".
SA3/MA1
2-1-A
2-2-A