H3-80.1
MAINTENANCE 3. TROUBLESHOOTING
006-252-05
06.30.01 FM3040
FM-DP L
Service Manual
H3-80.1
Error code
Error name
Significance
Probable cause and remedy
Detailed
information format
22E7
Sorter storage
inconsistency error
The sorter unit was performing storage
processing during initialization.
2-3-F
22EA
Storage bin number
detection failure
The storage response information from
the sorter unit and main body software
data do not correspond.
2-3-E2
2305
Possibility of no I/O
trace data
The I/O trace data storage file may not
exist.
The I/O trace is not started.
1. Set the I/O trace to ON using the M-Utility.
None
2602
DSP micro-program 1
load checksum warning
error
The transfer checksum value was
abnormal during DSP micro-program
load 1.
(Can be continued)
6-C
2603
DSP micro-program 2
load checksum warning
error
The transfer checksum value was
abnormal during DSP micro-program
load 2.
(Can be continued)
6-C
2604
DSP self-diagnosis
warning error
Results of the DSP initialization self-
diagnosis were abnormal.
6-C
Defect in the vicinity of the DSP of the IMG board. If no defects
are seen in particular, these corrections are not necessary.
1. Check that the CPU, IMG, input boards, MTH board related
connectors, image memory board are set to the MTH
board normally.
2. Replace the IMG board.
3. Replace the MTH board.
Defect in the vicinity of the DSP of the IMG board. If no defects
are seen in particular, these corrections are not necessary.
1. Check that the CPU, IMG, input boards, MTH board related
connectors, image memory board are set to the MTH
board normally.
2. Replace the IMG board.
3. Replace the MTH board.
Defect in the vicinity of the DSP of the IMG board. If no defects
are seen in particular, these corrections are not necessary.
1. Check that the CPU, IMG, input boards, MTH board related
connectors, image memory board are set to the MTH
board normally.
2. Replace the IMG board.
3. Replace the MTH board.
This error is most likely to have occurred due to other sorter
problems. Check for sorter problems before this error log.
•SIO transmission error (12C2)
•SIO reception error (12C3)
•Sorter unit detection command parameter error (12D6)
•Sorter unit detection invalid command (12D7)
•Sorter conveyance canceled (12E5)
This error is most likely to have occurred due to other sorter
problems. Check for sorter problems before this error log.
•SIO transmission error (12C2)
•SIO reception error (12C3)
•Sorter unit detection command parameter error (12D6)
•Sorter unit detection invalid command (12D7)
•Sorter conveyance canceled (12E5)