MAINTENANCE 3. TROUBLESHOOTING
Error code
Error name
Significance
Probable cause and remedy
Detailed
information format
1600
Uniformity correction
value write error
Failed to write the uniformity correction
value in the DSP external memory.
6-C
1601
Uniformity correction
value checksum
parameter write error
Failed to write the uniformity correction
value checksum data in the DSP externa
memory.
6-C
1602
Uniformity correction
value checksum error
Checksum of the uniformity correction
table is faulty.
6-C
1603
Uniformity correction
value checksum time-
out error
The checksum calculation of the DSP
uniformity correction value did not end
within the specified time (90 seconds).
6-C
Processes other than uniformity measurement can be continued
If image problems occur, re-do the uniformity measurement.
If the error occurs frequently, perform the following check.
Defect in the vicinity of the DSP of the IMG board.
1. Check that the CPU, IMG, input boards, MTH board related
connectors, image memory board are set to the MTH
board normally.
2. Replace the IMG board.
3. Replace the MTH board.
Processes other than uniformity measurement can be continued
If image problems occur, re-do the uniformity measurement.
If the error occurs frequently, perform the following check.
Defect in the vicinity of the DSP of the IMG board.
1. Check that the CPU, IMG, input boards, MTH board related
connectors, image memory board are set to the MTH
board normally.
2. Replace the IMG board.
3. Replace the MTH board.
Processes other than uniformity measurement can be continued
If image problems occur, re-do the uniformity measurement.
If the error occurs frequently, perform the following check.
Defect in the vicinity of the DSP of the IMG board.
1. Check that the CPU, IMG, input boards, MTH board related
connectors, image memory board are set to the MTH
board normally.
2. Replace the IMG board.
3. Replace the MTH board.
Processes other than uniformity measurement can be continued
If image problems occur, re-do the uniformity measurement.
If the error occurs frequently, perform the following check.
Defect in the vicinity of the DSP of the IMG board.
1. Check that the CPU, IMG, input boards, MTH board related
connectors, image memory board are set to the MTH
board normally.
2. Replace the IMG board.
3. Replace the MTH board.