H1-11
006-252-02
10.22.99 FM2569
FM-DP L
Service Manual
H1-11
MAINTENANCE 1. OUTLINE OF EQUIPMENT
Unit
Symbol
Name
Type
State at ON
Upper removal
SA1
Suction cup arm upper
Penetration type sensor Dead-point
unit [A]
dead-point detection sensor
(Pl:5 mm)
SA2
Suction cup arm turning point
Penetration type sensor Variable-point
detection sensor
(Pl:5 mm)
SA3
Film surface detection sensor
Penetration type sensor When suction unit is 0 to 6.5 mm
(Pl:5 mm)
above film
SA4
Film trailing-edge detection
Reflection type sensor
When film is in removal unit
sensor
conveyance roller
SA5
Film pack/shutter detection
Penetration type sensor When film pack/shutter is detected
sensor
(Pl:5 mm)
SA6
Tray detection switch
Micro switch
When tray is pushed in
SA7
Tray lock detection sensor
Penetration type sensor When lock is released
(Pl:5 mm)
SA8
Suction detection sensor
Pressure sensor
When film is sucked in
Lower removal
SB1
Suction cup arm upper
Penetration type sensor Dead-point
unit [B]
dead-point detection sensor
(Pl:5 mm)
(Optional)
SB2
Suction cup arm turning point
Penetration type sensor Variable-point
detection sensor
(Pl:5 mm)
SB3
Film surface detection sensor
Penetration type sensor When suction unit is 0 to 6.5 mm
(Pl:5 mm)
above film
SB4
Film trailing-edge detection
Reflection type sensor
When film is in removal unit
sensor
conveyance roller
SB5
Film pack/shutter detection
Penetration type sensor When film pack/shutter is detected
sensor
(Pl:5 mm)
SB6
Tray detection switch
Micro switch
When tray is pushed in
SB7
Tray lock detection sensor
Penetration type sensor When lock is released
(Pl:5 mm)
TP3F0201.EPS
<SUPPLEMENTARY NOTE>
The ON of each sensor and switch means the following;
• Penetration type sensor: Light-tight state
• Reflection type sensor: Film present
• Microswitch: Power conducted
• Pressure sensor: Film sucked