4 - 2
Set Reference
Connect OMNIS
CANNER
and OMNIR
EMOTE
through
their test connector ports using the Link Adapter
and the Channel Adapter. Select
Set Reference
to display the
SET REFERENCE
screen.
OMNIS
CANNER
will immediately start to communi-
cate with OMNIR
EMOTE
.
While the reference data is being established, a
gauge indicates the progress.
OMNIS
CANNER
notifies you when the process has been completed. The
REMOTE INFO
screen will be displayed.
Note
Set Reference is needed to periodically verify and establish the signal
reference for attenuation.
Press ESC to return to the
SETUP
screen.
Set Reference with the OMNIF
IBER
Adapters
Note
The fiber reference expires each day at midnight. It must be set each
calendar day that OMNIF
IBER
adapters are used. The reference value
should also be set every time the test configuration changes.
A reference value must be established before a
loss measurement can be calculated.
With the OMNIF
IBER
adapters attached to
OMNIS
CANNER
and OMNIR
EMOTE
, and the launch
cable(s) (jumper(s)) in place, select
Set Reference
to display the
CONNECT JUMPERS
screen.
Connect the jumpers as displayed and press
to continue.
The
SET REFERENCE
screen will be displayed
while the reference data is being established.
OMNIS
CANNER
notifies you when the process has
been completed and the
FIBER INFO
screen will
be displayed.
Summary of Contents for OMNIFiber
Page 1: ...OMNIScanner 2 OMNIFiber ...
Page 9: ...OMNISCANNER User Guide TOC 6 ...
Page 25: ...OMNISCANNER User Guide 1 16 ...
Page 33: ...OMNISCANNER User Guide 2 8 ...
Page 41: ...OMNISCANNER User Guide 2 16 ...
Page 103: ...OMNISCANNER User Guide 7 6 ...
Page 111: ...OMNISCANNER User Guide 9 2 ...
Page 129: ...OMNISCANNER User Guide A 8 ...
Page 133: ...OMNISCANNER User Guide C 2 ...
Page 136: ......