2 - 3
OMNIS
CANNER
User Guide
• Setup Projects
• View Graphs
• Edit Autotest information
The following sections provide detail on the certification process.
Setting the Reference
A reference must be set when OMNIS
CANNER
and OMNIR
EMOTE
are used
as a set for the first time. After the initial reference setting, the process
should be repeated periodically to ensure unit verification.
OMNIS
CANNER
stores reference data for up to 5 remote units by serial
number.
1. To set the reference, connect OMNIS
CANNER
and OMNIR
EMOTE
using the PERMANENT LINK
ADAPTER and CHANNEL ADAPTER supplied
with the scanner (as shown).
2. Turn OMNIS
CANNER
on.
3. Press
(Setup).
4. Select
Set Reference
.
1
-/.
abc
def
ghi
jkl
mno
prs
tuv
wxy
qz
2
3
4
5
6
7
8
9
0
?
A/a
AUTOTEST
ESC
ENTER
CAT 6
Channel
Adapter
CAT 6
Perminant
Link
Adapter