5730A
Operators Manual
4-28
9.
The display now shows the amplitude of the entry. If an entry error is made, push
to clear the display, then reenter the value. If the most recent digit entered is in error,
push
to clear that digit. The illustration of the display below assumes an entry of
20 dBm:
dBm
kHz
1.0000
42.00000
+/- 1.37 % (1 year) 213 days since calibration
4/14/2013 13:40
Setup
Menu
Standby
μV
kHz
1.0000
Auto Range 1.1 mV
350.00
Wideband
ON
20dBm
hhp125.eps
10.
With the numeric keypad, enter a frequency, followed by
or
if necessary.
The display now shows the amplitude and frequency of the entry. If an entry error is
made, push
to clear the display, then reenter the value. If the most recent digit
entered is in error, push
to clear that digit. The display shown below assumes an
entry of 21 MHz:
dBm
kHz
1.0000
42.00000
+/- 1.37 % (1 year) 213 days since calibration
4/18/2013 13:40
Setup
Menu
Standby
μV
kHz
1.0000
Auto Range 1.1 mV
350.00
Wideband
ON
20dBm 21MHz
hhp126.eps
Summary of Contents for 5730A
Page 4: ......
Page 10: ...5730A Operators Manual x ...
Page 12: ...5730A Operators Manual xii ...
Page 14: ...5730A Operators Manual xiv ...
Page 16: ...5730A Operators Manual 1 2 ...
Page 17: ...Introduction and Specifications Introduction 1 1 3 ...
Page 18: ......
Page 52: ...5730A Operators Manual 2 2 ...
Page 58: ...5730A Operators Manual 3 2 ...
Page 124: ...5730A Operators Manual 5 2 ...
Page 136: ...5730A Operators Manual 6 2 ...
Page 216: ...5730A Operators Manual 7 24 ...
Page 218: ...5730A Operators Manual 8 2 ...
Page 222: ...5730A Operators Manual 8 6 ...
Page 248: ...5730A Operators Manual A 26 ...
Page 249: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...