5730A
Operators Manual
4-10
UUT
UUT
Calibrator
Calibrator
Note:
Keep the SENSE leads as short as possible.
Be careful not to exceed the capacitive load limit of 1000 pF up to
220 V, 600 pF 220 V to 1100 V. (1000 pF with 5725A Amplifier.)
INPUT
SENSE
4-WIRE
W
HI
LO
A
HI
LO
GUARD
INPUT
SENSE
4-WIRE
W
HI
LO
A
HI
LO
GUARD
Triaxial Cable
Triaxial Cable
Triaxial Cable
External
Sense
ON
External
Guard
OFF
External
Sense
OFF
External
Guard
OFF
A
B
hhp012.eps
Figure 4-2. UUT Connections: AC Voltage >10 kHz