Front-Panel Operation
Linearity Check with Offset and Scale
4
4-43
Note
The message “Error >+999.9999%” occurs because in this case the
reference voltage is 0.
3.
Touch
Offset
to identify this as the DMM zero-scale endpoint. The display changes
to:
dBm
kHz
1.0000
42.00000
+/- 3% (1 year) 213 days since calibration
12/21/12 10:52am
External
Sense
OFF
External
Guard
OFF
Wideband
OFF
Setup
Menu
Standby
mV
Auto Range 220 mV
- 1.3000
Offset
Scale
External
Sense
OFF
Reference
0.00000 mV
Offset
1.30000 mV
hhp137.eps
4.
Use the numeric keypad to set the Calibrator to a point near the end
of the DMM
range. (This example uses 19.9 V.) The displays
change to:
dBm
kHz
1.0000
42.00000
+/- 4.3 ppm (1 year) 213 days since calibration
4/18/2013 14:33
External
Sense
OFF
External
Guard
OFF
Wideband
OFF
Setup
Menu
Standby
V
Auto Range 22 V
19.898700
Offset
Scale
External
Sense
OFF
Reference
19.9 V
Offset
0.001300 V
hhp138.eps
Summary of Contents for 5730A
Page 4: ......
Page 10: ...5730A Operators Manual x ...
Page 12: ...5730A Operators Manual xii ...
Page 14: ...5730A Operators Manual xiv ...
Page 16: ...5730A Operators Manual 1 2 ...
Page 17: ...Introduction and Specifications Introduction 1 1 3 ...
Page 18: ......
Page 52: ...5730A Operators Manual 2 2 ...
Page 58: ...5730A Operators Manual 3 2 ...
Page 124: ...5730A Operators Manual 5 2 ...
Page 136: ...5730A Operators Manual 6 2 ...
Page 216: ...5730A Operators Manual 7 24 ...
Page 218: ...5730A Operators Manual 8 2 ...
Page 222: ...5730A Operators Manual 8 6 ...
Page 248: ...5730A Operators Manual A 26 ...
Page 249: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...