Front-Panel Operation
Set the Output
4
4-21
In dc current output, the selections available at the bottom are:
•
Current Output
•
External Guard
•
Boost
•
Wideband
•
Setup Menu
In addition, the Offset and Scale functions are available. The output range can be locked
in dc current.
Current Output
selects one of three locations for non-boosted (current sourced by the
Calibrator, for example) current output:
•
OUTPUT binding post (NORMAL, which is the OUTPUT HI binding post, also the
default setting)
•
AUX, which is the AUX CURRENT OUTPUT binding post
•
5725A which is the 5725A Amplifier binding posts. (The 5725A must be turned on,
but not necessarily activated.)
Non-boosted current from the Calibrator is not available at the 52120A amplifier binding
posts.