5730A
Operators Manual
4-22
AC Current Output
To set an ac current output:
1.
Make sure the Calibrator is in standby (STANDBY annunciator lit). Push
if
necessary.
2.
If the UUT is not connected, connect it now as described earlier in this chapter under
“Connect the Calibrator to a UUT.”
3.
Set the UUT to measure ac current on the appropriate range.
4.
Enter a current magnitude with the numeric keypad.
5.
Push
or
if necessary.
6.
Push
.
7.
The display now shows the amplitude of the entry. If an entry error is made, push
to clear the display, then reenter the value. If the most recent digit entered is in error,
push
to clear that digit. The illustration of the display below assumes an entry of
125 mA:
+/- 0.00050mV (1 year) 213 days since calibration
3/14/2013 10:52am
External
Sense
OFF
External
Guard
OFF
Wideband
OFF
Setup
Menu
Operate
mV
0.00000
Auto Range 220 mV
125mA
External
Sense
OFF
hhp119.eps
Summary of Contents for 5730A
Page 4: ......
Page 10: ...5730A Operators Manual x ...
Page 12: ...5730A Operators Manual xii ...
Page 14: ...5730A Operators Manual xiv ...
Page 16: ...5730A Operators Manual 1 2 ...
Page 17: ...Introduction and Specifications Introduction 1 1 3 ...
Page 18: ......
Page 52: ...5730A Operators Manual 2 2 ...
Page 58: ...5730A Operators Manual 3 2 ...
Page 124: ...5730A Operators Manual 5 2 ...
Page 136: ...5730A Operators Manual 6 2 ...
Page 216: ...5730A Operators Manual 7 24 ...
Page 218: ...5730A Operators Manual 8 2 ...
Page 222: ...5730A Operators Manual 8 6 ...
Page 248: ...5730A Operators Manual A 26 ...
Page 249: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...