5730A
Operators Manual
7-16
3.
Touch
220 V
and this screen is shown:
220 V
Enter nominal value then press ENTER.
Cancel
Range
Adjustment
Setup Menu
Calibration
hhp216.eps
5.
Connect the 732B, 8508A (Null Detector), and 752B in a 10:1 configuration, as
shown in Figure 7-6.
6.
Multiply the 732B value by 10, and enter this new value. The Calibrator output will
be nulled to this new value.
7.
Push
to show this display:
220 V
Range
Adjustment
Setup Menu
Calibration
External
Sense
OFF
External
Guard
OFF
Continue
V
200.00000
External
Sense
OFF
Reference
200.000000 dBm
Error
= +0.00ppm
Standby
Press OPERATE. Adjust with knob then press Continue.
hhp217.eps
8.
Push
to activate the Calibrator output. Turn the output adjustment knob on the
Calibrator until a null is achieved on the null detector, and then touch
Continue
.
9.
Touch
Save
to be prompted to enter the security passcode if the Calibrator has not
been unsecured already. It then saves the new range adjustment for the 220 V dc
range.
10.
If
Discard
is touched, the result is discarded and the Range Calibration screen is
shown.
Summary of Contents for 5730A
Page 4: ......
Page 10: ...5730A Operators Manual x ...
Page 12: ...5730A Operators Manual xii ...
Page 14: ...5730A Operators Manual xiv ...
Page 16: ...5730A Operators Manual 1 2 ...
Page 17: ...Introduction and Specifications Introduction 1 1 3 ...
Page 18: ......
Page 52: ...5730A Operators Manual 2 2 ...
Page 58: ...5730A Operators Manual 3 2 ...
Page 124: ...5730A Operators Manual 5 2 ...
Page 136: ...5730A Operators Manual 6 2 ...
Page 216: ...5730A Operators Manual 7 24 ...
Page 218: ...5730A Operators Manual 8 2 ...
Page 222: ...5730A Operators Manual 8 6 ...
Page 248: ...5730A Operators Manual A 26 ...
Page 249: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...