5730A
Operators Manual
4-56
The second method is called “true value”. Both methods can be used in this Calibrator.
The nominal method of error calculation uses the formula:
reference value – edit value
reference value
The nominal method is useful to check the error of the Calibrator itself, when its
performance is verified against a more accurate measuring device.
The true value method of error calculation uses the formula:
Reference value – edit value
edit value
With either the nominal or true value method, small changes in output value result in a
calculated error that is the same. In the example above, the display will show the error as
-30.0 ppm.
The true value method is useful for large changes in output value. For example, if
10.0000 V is applied to an analog meter, and then the Calibrator output is adjusted to
11.0000 V so that the analog meter now reads exactly 10 V, the true value method will
show:
nominal= +10.0000 V
rel err= -9.0909 %
-9.0909 % represents the relative error of the analog meter when compared to the true
value (11.0000 V in this case).
To select the UUT error calculation method:
1.
Touch
Setup>Instrument Setup
.
2.
Touch
Error Reference
until it shows the desired choice. The choices are "Nominal
Value" or "True Value."
Summary of Contents for 5730A
Page 4: ......
Page 10: ...5730A Operators Manual x ...
Page 12: ...5730A Operators Manual xii ...
Page 14: ...5730A Operators Manual xiv ...
Page 16: ...5730A Operators Manual 1 2 ...
Page 17: ...Introduction and Specifications Introduction 1 1 3 ...
Page 18: ......
Page 52: ...5730A Operators Manual 2 2 ...
Page 58: ...5730A Operators Manual 3 2 ...
Page 124: ...5730A Operators Manual 5 2 ...
Page 136: ...5730A Operators Manual 6 2 ...
Page 216: ...5730A Operators Manual 7 24 ...
Page 218: ...5730A Operators Manual 8 2 ...
Page 222: ...5730A Operators Manual 8 6 ...
Page 248: ...5730A Operators Manual A 26 ...
Page 249: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...