5730A
Operators Manual
4-38
The example in Figure 4-8 assumes no offset error. The formula for scale error yields:
Scale Error=
19.9 -19.903
= -0.000151= -0.0151 %
19.9
Calibrator Setting
0
Meter Reading
19.903 V
19.9 V
Scale
Error =
-3 mV
19.9 V
Ideal
Actual
hhp018.eps
Figure 4-8. Scale Error
Linearity Error
Linearity error occurs when the response curve of a meter deviates from a straight line.
This type of error is measured by fixing zero and near-full-scale endpoints on the
response curve, drawing a line through the points, then measuring how far the meter
response deviates from the straight line at various points. The error value is computed
relative to the selected full-scale point. The formula for linearity error is:
Linearity Error=
(Nominal setting) – (Adjusted 5730A Setting)
(Nominal Full Scale)
Where the “nominal setting” is calculated by subtracting the offset error and a
proportional part of the scale error.
Summary of Contents for 5730A
Page 4: ......
Page 10: ...5730A Operators Manual x ...
Page 12: ...5730A Operators Manual xii ...
Page 14: ...5730A Operators Manual xiv ...
Page 16: ...5730A Operators Manual 1 2 ...
Page 17: ...Introduction and Specifications Introduction 1 1 3 ...
Page 18: ......
Page 52: ...5730A Operators Manual 2 2 ...
Page 58: ...5730A Operators Manual 3 2 ...
Page 124: ...5730A Operators Manual 5 2 ...
Page 136: ...5730A Operators Manual 6 2 ...
Page 216: ...5730A Operators Manual 7 24 ...
Page 218: ...5730A Operators Manual 8 2 ...
Page 222: ...5730A Operators Manual 8 6 ...
Page 248: ...5730A Operators Manual A 26 ...
Page 249: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...