Introduction
and
Specifications
Electrical Specifications
1
1-35
52120A/COIL 3 KA 25-Turn Coil
Number of turns
.................................................... 25
Minimum internal jaw dimension to clear wires
26 mm (width) x 36 mm (length)
Maximum input current
........................................ 120 A continuous with built-in 12 V fan on
Maximum voltage
.................................................. 4.5 V rms
Accuracy
Input Current
[1]
Frequency
Effective Current
Amp-turns
Coil Accuracy
[2]
±
(% of Amp-turns + % of 52120A range)
% of Amp-turns
% of 52120A Range
0 A to 100 A
DC
0 to 2500
0.7 %
0.7 %
0 A to 120 A
10 Hz to 65 Hz
0 to 3000
0.7 %
0.7 %
0 A to 120 A
65 Hz to 300 Hz
0 to 3000
0.7 %
0.7 %
0 A to 40 A
300 Hz to 1 kHz
0 to 1000
0.7 %
0.7 %
0 A to 12 A
1 kHz to 3 kHz
0 to 300
0.8 %
1.0 %
0 A to 3 A
3 kHz to 6 kHz
0 to 75
1.5 %
1.0 %
0 A to 1 A
6 kHz to 10 kHz
0 to 25
5.0 %
1.0 %
Notes:
1.
The inductance and mutual inductance of the 25 turn coil and clamp that is measured causes a frequency dependent compliance
voltage across the coil. The length and configuration of the cables that connect the current to the coil also have an effect.
Maximum input current is 120 A input at approximately 100 Hz. Maximum current input decreases to approximately 0.8 A at
10 kHz.
2.
Includes coil/clamp interaction.
52120A/COIL 6 KA 50-Turn Coil
Number of turns
.................................................... 50
Minimum flexible probe length
............................ 500 mm
Maximum input current
........................................ 120 A continuous with built-in 12 V fan on
Maximum voltage
.................................................. 4.5 V rms
Accuracy
Input Current
[1]
Frequency
Effective Current
Amp-turns
Coil Accuracy
[2]
±
(% of Amp-turns + % of 52120A range)
% of Amp-turns
% of 52120A Range
0 A to 100 A
DC
0 to 5000
0.7 %
0.7 %
0 A to 120 A
10 Hz to 65 Hz
0 to 6000
0.7 %
0.7 %
0 A to 120 A
65 Hz to 300 Hz
0 to 6000
0.7 %
0.7 %
0 A to 120 A
300 Hz to 1 kHz
0 to 6000
0.7 %
0.7 %
0 A to 120 A
1 kHz to 3 kHz
0 to 6000
0.8 %
1.0 %
0 A to 25 A
3 kHz to 6 kHz
0 to 1250
1.5 %
1.0 %
0 A to 13 A
6 kHz to 10 kHz
0 to 650
5.0 %
1.0 %
Notes:
1.
The inductance and mutual inductance of the 50 turn coil causes a frequency dependent compliance voltage across the coil.
Maximum frequency for 120 A input current is approximately 600 Hz. Maximum current input decreases to approximately 13 A at
10 kHz.
2.
Includes coil/probe interaction.
Note
The specifications for these coils are at 99 % confidence level and are the combined
accuracy of the coil and a 52120A. If the coils are used with other current sources
the calibration accuracy of the coils alone is 0.65 % (99 % confidence level) from
0 Hz to 10 kHz.
Summary of Contents for 5730A
Page 4: ......
Page 10: ...5730A Operators Manual x ...
Page 12: ...5730A Operators Manual xii ...
Page 14: ...5730A Operators Manual xiv ...
Page 16: ...5730A Operators Manual 1 2 ...
Page 17: ...Introduction and Specifications Introduction 1 1 3 ...
Page 18: ......
Page 52: ...5730A Operators Manual 2 2 ...
Page 58: ...5730A Operators Manual 3 2 ...
Page 124: ...5730A Operators Manual 5 2 ...
Page 136: ...5730A Operators Manual 6 2 ...
Page 216: ...5730A Operators Manual 7 24 ...
Page 218: ...5730A Operators Manual 8 2 ...
Page 222: ...5730A Operators Manual 8 6 ...
Page 248: ...5730A Operators Manual A 26 ...
Page 249: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...