Remote Commands and Syntax
Commands
6
6-45
Table 6-5.
Commands (cont.)
SP_SET?
Description
Sequential command.
Returns the serial port settings contained in nonvolatile memory.
Response
1. (Integer) One of these baud rates: 9600, 19200, 38400, 57600 or 115200
2. (String) TERM or COMP
(Response type)
3. (String) XON, RTS, or NOSTALL
(Stall method)
4. (String) DBIT7 or DBIT8
(Data bits)
5. (String) SBIT1 or SBIT2
(Stop bits)
6. (String) PNONE, PEVEN, PODD, or PIGNORE (Parity)
7. (String) CR, LF, or CRLF
(End-Of-Line)
Example 9600,TERM,XON,DBIT8,SBIT1,PNONE,CRLF
*SRE
Description
Sequential command.
Loads a byte into the Service Request Enable register (SRE), described under "Check
the Calibrator Status."
Parameter
The decimal equivalent of the binary number to load into the register.
Example
*SRE 56
Enables bits 3 (IIR), 4 (MAV), and 5 (ESR) in the Service Request Enable register.
*SRE?
Description
Sequential command.
Returns the byte from the Service Request Enable register, described under "Check the
Calibrator Status."
Parameter None
Response
(Integer) The decimal equivalent of the register byte.
Example
*SRE?
Returns: "56" if bits 3 (IIR), 4 (MAV), and 5 (ESR) are enabled (1) and the rest of the bits
are disabled (0). (See "Check the Calibrator Status" for details.)
Summary of Contents for 5730A
Page 4: ......
Page 10: ...5730A Operators Manual x ...
Page 12: ...5730A Operators Manual xii ...
Page 14: ...5730A Operators Manual xiv ...
Page 16: ...5730A Operators Manual 1 2 ...
Page 17: ...Introduction and Specifications Introduction 1 1 3 ...
Page 18: ......
Page 52: ...5730A Operators Manual 2 2 ...
Page 58: ...5730A Operators Manual 3 2 ...
Page 124: ...5730A Operators Manual 5 2 ...
Page 136: ...5730A Operators Manual 6 2 ...
Page 216: ...5730A Operators Manual 7 24 ...
Page 218: ...5730A Operators Manual 8 2 ...
Page 222: ...5730A Operators Manual 8 6 ...
Page 248: ...5730A Operators Manual A 26 ...
Page 249: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...