Front-Panel Operation
Linearity Check with Offset and Scale
4
4-45
The 10.000208 V output setting is calculated by the Calibrator with the subsequent
equation:
1 -
19.9 V – 19.903 V
= 1.0001508
19.9 V
This is applied to 10 V to yields:
10 V x 1.0001508 = 10.001508 V
The 1.3 mV zero offset is then subtracted:
10.001508 V
−
0.0013 V = 10.000208 V
8.
Use the output adjustment controls to adjust the Calibrator
output for a reading of
10.0 V (the reference value) on the DMM.
The displays change to:
dBm
kHz
1.0000
42.00000
+/- 4.3 ppm (1 year) 213 days since calibration
4/18/2013 14:33
External
Sense
OFF
External
Guard
OFF
Wideband
OFF
Setup
Menu
Standby
V
Auto Range 11 V
9.99
3
208
Offset
Scale
Reference
10.000000 V
Offset
0.001300 V
Linearity Error
= +0.0352%
New Reference
hhp141.eps
The display now shows that the DMM scale error at 19.9 V is -0.0151 %, and the DMM
linearity error at 10 V is +0.0352 %.
Summary of Contents for 5730A
Page 4: ......
Page 10: ...5730A Operators Manual x ...
Page 12: ...5730A Operators Manual xii ...
Page 14: ...5730A Operators Manual xiv ...
Page 16: ...5730A Operators Manual 1 2 ...
Page 17: ...Introduction and Specifications Introduction 1 1 3 ...
Page 18: ......
Page 52: ...5730A Operators Manual 2 2 ...
Page 58: ...5730A Operators Manual 3 2 ...
Page 124: ...5730A Operators Manual 5 2 ...
Page 136: ...5730A Operators Manual 6 2 ...
Page 216: ...5730A Operators Manual 7 24 ...
Page 218: ...5730A Operators Manual 8 2 ...
Page 222: ...5730A Operators Manual 8 6 ...
Page 248: ...5730A Operators Manual A 26 ...
Page 249: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...