5730A
Operators Manual
4-44
5.
Use the output adjustment controls to adjust the Calibrator
output for a reading of
19.9 V (the reference value) on the DMM.
The displays change to:
dBm
kHz
1.0000
42.00000
+/- 4.3 ppm (1 year) 213 days since calibration
4/18/2013 14:33
External
Sense
OFF
External
Guard
OFF
Wideband
OFF
Setup
Menu
Standby
V
Auto Range 22 V
19.90
1
700
Offset
Scale
Reference
19.900000 V
Offset
0.001300 V
Error
= -0.0151%
New Reference
hhp139.eps
6.
Touch
Scale
to identify this as the full-scale endpoint for
the DMM range.
7.
To check the linearity error of the DMM, select an output at a
linearity checkpoint
somewhere in the middle of the range, such
as 10 V. If 10 V is entered, this changes
the display to:
dBm
kHz
1.0000
42.00000
+/- 4.3 ppm (1 year) 213 days since calibration
4/18/2013 14:33
External
Sense
OFF
External
Guard
OFF
Wideband
OFF
Setup
Menu
Standby
V
Auto Range 10 V
10.000208
Offset
Scale
Reference
10.000000 V
Offset
0.001300 V
Scale Error
= -0.0151%
hhp140.eps
Summary of Contents for 5730A
Page 4: ......
Page 10: ...5730A Operators Manual x ...
Page 12: ...5730A Operators Manual xii ...
Page 14: ...5730A Operators Manual xiv ...
Page 16: ...5730A Operators Manual 1 2 ...
Page 17: ...Introduction and Specifications Introduction 1 1 3 ...
Page 18: ......
Page 52: ...5730A Operators Manual 2 2 ...
Page 58: ...5730A Operators Manual 3 2 ...
Page 124: ...5730A Operators Manual 5 2 ...
Page 136: ...5730A Operators Manual 6 2 ...
Page 216: ...5730A Operators Manual 7 24 ...
Page 218: ...5730A Operators Manual 8 2 ...
Page 222: ...5730A Operators Manual 8 6 ...
Page 248: ...5730A Operators Manual A 26 ...
Page 249: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...