Remote Commands and Syntax
Commands
6
6-35
Table 6-5.
Commands (cont.)
ISCE
Description
Sequential command.
Loads a byte into the Instrument Status Change Enable register described under "Check
the Calibrator Status."
Parameter
The decimal equivalent of the binary number to load into the register.
Example
ISCE 56
Enables bits 3 (BOOST), 4 (RCOMP), and 5 (RLOCK) in the Service Request Enable
register.
ISCE?
Description
Sequential command.
Returns the byte from the Instrument Status Change Enable register, described under
"Check the Calibrator Status."
Parameter None
Response
The decimal equivalent of the register contents byte.
Example
ISCE?
Returns: "4" if bit 3 (BOOST) is enabled (1) and the rest of the bits are disabled (0). (See
"Check the Calibrator Status" for details.)
ISCR?
Description
Sequential command.
Returns and clears the byte from the Instrument Status Change Register, described
under "Check the Calibrator Status."
Parameter None
Response
The decimal equivalent of the register contents byte.
Example
ISCR?
Returns: "8" if bit 3 (BOOST) is set (1) and the rest of the bits are reset (0). (See "Check
the Calibrator Status" for details.)
ISR?
Description
Sequential command.
Returns and clears the byte from the Instrument Status Register, described under
"Check the Calibrator Status."
Parameter None
Response
The decimal equivalent of the register contents byte.
Example
ISR?
Returns: "16" if bit 4 (RLOCK) is set (1) and the rest of the bits are reset (0). (See "Check
the Calibrator Status" for details.)
LCOMP_52120
Description
Overlapped command,. Ignored if not in remote. Sets inductance compensation for
attached 52120As.
Parameter
ON (enables LCOMP for all attached 52120s)
OFF (disables LCOMP for all attached 52120s)
Summary of Contents for 5730A
Page 4: ......
Page 10: ...5730A Operators Manual x ...
Page 12: ...5730A Operators Manual xii ...
Page 14: ...5730A Operators Manual xiv ...
Page 16: ...5730A Operators Manual 1 2 ...
Page 17: ...Introduction and Specifications Introduction 1 1 3 ...
Page 18: ......
Page 52: ...5730A Operators Manual 2 2 ...
Page 58: ...5730A Operators Manual 3 2 ...
Page 124: ...5730A Operators Manual 5 2 ...
Page 136: ...5730A Operators Manual 6 2 ...
Page 216: ...5730A Operators Manual 7 24 ...
Page 218: ...5730A Operators Manual 8 2 ...
Page 222: ...5730A Operators Manual 8 6 ...
Page 248: ...5730A Operators Manual A 26 ...
Page 249: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...