Operator's Manual
|
3-12
The
▲
Paw (High Pressure) alarm is violated
Three seconds have elapsed.
Manual Inflation is only available in A/CMV and SIMV modes.
Manual Inflation may be prematurely cycled off in the first
several breaths in Pressure Control when the initial flow has
not yet been optimized.
3.2.14 Humidifier On Button
Figure 15 – Humidifier On Button
Range:
19ºC to 39ºC
This touch button activates the built-in humidifier. Pressing this button displays
the set target temperature in the adjacent window. While the display is
blinking, use the
▲
Up/
▼
Down
controls to adjust the target temperature.
When temperature adjustment is complete and the following occurs:
Five seconds have elapsed without touching the control
The Humidifier On button is tapped again
Another parameter is selected for adjustment
The display stops blinking and the measured temperature is displayed.
While the humidifier is On, the target temperature can be readjusted by
pressing the
Humidifier On
button and using the
▲
Up/
▼
Down
controls.
Preheating the humidifier for 30 minutes prior to beginning
ventilation will improve the heating performance of the
humidifier.
During ventilation (On condition) the displayed temperature is the measured
temperature at the patient connector. In the Setting condition, the displayed
temperature is the measured temperature at the humidifier bottle outlet.
To turn the humidifier Off, press and hold the
Humidifier On
button for three
Humidifier On Button page 3-12 for more details.