Operator's Manual
|
5-5
Remove the plastic packaging from the humidifier bottle and
heat sink before use.
Figure 50 – Humidifier Bottle and Heat Sink
1. Ensure that the patient circuit and humidifier are assembled correctly.
2. Open the humidifier clamp knob located on the upper case of the
ventilator.
3. Remove the humidifier bottle from the ventilator.
4. Open the humidifier bottle, remove the heat sink and place absorbent
paper between the heat sink ribs.
5. Position the heat sink inside the lower half of the bottle.
6. Fill the lower half with sterile, distilled water, up to the FULL line. Attach
the upper half of the bottle to the lower half.
7. Re-attach the humidifier bottle to the ventilator. Secure the clamp knob.
8. Connect the short side of the Humidifier Temperature Probe cable to the
port on the top of the humidifier bottle.
9. Plug the electrical connector into the side of the FLIGHT 50 Ventilator and
then connect the other end of the cable to the temperature probe port on
the exhalation valve.
5.5.3 Set Up and Operation
1. From the Off/Standby condition, switch the ventilator to the Setting
condition by pressing
On/Standby
once.
2. Turn the humidifier on by pressing the
Humidifier On
button.
3. Adjust humidification temperature.