Operator's Manual
|
5-3
Be careful to position the cord so that it won’t interfere
with the Air/Oxygen Entrainment Mixer attachment.
Figure 48 – Power Cord Ferrite
6. Verify that the A.C. power cord supplied with the FLIGHT 50 Ventilator
has the Power Cord Ferrite attached between the adapter box and the
ventilator.
7. If utilizing external power, plug the ventilator’s electric cord into a
properly grounded outlet.
8. If using internal battery, ensure that battery is fully charged.
Always use the A.C. power cord supplied with the FLIGHT
50 Ventilator that has the Power Cord Ferrite attached
between the adapter box and the ventilator to ensure that
the FLIGHT 50 Ventilator meets EMC requirements.
9. Attach a patient circuit as follows:
For use without a humidifier, attach one end of the 22 mm ID
breathing circuit to the Gas Output on the FLIGHT 50 Ventilator.
When using a humidifier, connect one end of the short humidifier
tubing to the Gas Output on the FLIGHT 50 Ventilator and the other
end to the inlet port of the humidifier. Then attach one end of the
22 mm ID breathing circuit limb to the outlet
port of the humidifier.
Attach the Proximal Inline Filter with tubing to the Airway Pressure
connection located below the Gas Output port.
Attach one end of the Proximal Pressure sensing line to the Proximal
Inline Filter as shown in Figure 49, Detail A.
For use without a humidifier or HME, attach the other end of the
Proximal Pressure line to the pressure port on the exhalation valve.
See Detail B, Figure 5–1.
For use with an HME, attach the Proximal Pressure line to a
pressure tee adapter (p/n NP130-22) on the patient side of the
HME. Use the cap that comes with the tee adapter to seal the
pressure port on the exhalation valve.
For use with a humidifier, attach the Proximal Pressure line to a
pressure tee adapter (p/n NP130-22) before the inlet port on the