Operator's Manual
|
3-34
HOME
CAREGIVERS: The unit of measure, along with other
parameter settings, should be established by the patient’s
physician or Homecare Dealer.
3.8.1.3
Set Up
The Set Up parameter allows the operator to set the ventilator’s start up
settings. There are two selections, DEFAULT or USER. Use the
▲
Up/
▼
Down
button to toggle between selections.
When USER is selected, all of the ventilation parameters in effect at
shutdown will be saved. The saved parameters will appear next time the
ventilator is powered on.
When DEFAULT is selected, factory set default parameters will appear next
time the ventilator is powered on. Default parameters are listed on page
2-1.
HOME CAREGIVERS: This parameter should always be set to
USER for home use to ensure that when the FLIGHT 50
Ventilator is powered off and on the physician directed
ventilation settings remain in place.
3.8.1.4
Auto Panel Lock
This setting allows the user to enable (turn on) or disable (turn off) the Auto
Panel Lock feature. For a complete description of Auto Panel Lock see page
3-34. The Auto Panel Lock feature is factory set at Disabled or off. Use the
▲
Up/
▼
Down
buttons to toggle between Enabled and Disabled.
HOME CAREGIVERS: FLIGHT MEDICAL recommends that the
Auto Panel Lock feature be enabled in homecare
environments as an added safety feature to prevent
accidental changes to panel controls.
3.8.1.5
Exiting User Set Up
Exit by pressing the
Silence/Reset
button at any time.