Operator's Manual
|
7-3
7.2.2 Alarms and Indicators Check
1. If the FLIGHT 50 with a humidifier is being used, press
Humidifier On
button and use
▲
/
▼
control button to set the desired temperature.
2. Press the
On/Standby
button again to exit Settings condition and start
ventilation.
7.2.2.1
High
▲Paw Alarm
To check the High
▲
Paw alarm:
1. Set the High Paw alarm limit to 20 cmH
2
O/mbar
2. Verify that an audible and visual High Paw alarm occurs and that
inspiration ends when pressure reaches the high limit.
3. Set the High Paw alarm limit back to 99 cmH
2
O/97 mbar.
4. Verify that High Pressure alarm is deactivated and the
▲
Paw LED latches.
5. Press the
Silence/Reset
button to clear the alarm.
To check the Low
▼
Paw alarm:
1. Disconnect the test lung from the breathing circuit and verify that after
two breaths an audible and visual Low Paw alarm occurs.
2. Attach the test lung to the breathing circuit.
3. Verify that the audible alarm ceases and the
▼
Paw LED latches.
4. Press the
Silence/Reset
button to clear the alarm and the message in the
Message Display Window.
To check the Humidifier alarm (on V51-00000-60 model only):
1. Read the temperature displayed in the humidifier temperature window.
2. Press the
Humidifier On
button once and use the
▼
Down control to lower
the set humidifier temperature by 5º.
3. Verify that an audible alarm sounds and the Message Display Window
reads “
High Prox Temp
” and the humidifier is powered off.
4. Press the
Silence/Reset
button to clear the alarm.
5. Press the
Humidifier On
button and adjust the set temperature back to
the desired temperature.
6. Unclamp the humidifier bottle and remove it from the heating element.
7. Verify that an audible alarm sounds and Message Display Window reads
“
Humidifier Fail
” and the humidifier is powered off.
8. Press the
Silence/Reset
button to clear the alarm.
9. Reconnect the humidifier bottle.
10. Press the
Humidifier On
button.
11. Disconnect the temperature probe from the side of the FLIGHT 50.
12. Verify that an audible alarm sounds and the Message Display Window
reads “
Check Temp Probe
” and the humidifier is powered off.
13. Press the
Silence/Reset
button to clear the alarm.