Specifications
SYMBOLS and SPECIFICATIONS
Operator's Manual
|
2-3
Controls/Alarms/Monitors
Range/Selection
MODE (Pressure or Volume Control)
A/CMV
SIMV
SPONT
Volume Control (Tidal Volume)
100 to 2,200 mL, ATPS, ± 10%
Pressure Control (Target Pressure)
PEEP +5 to 60 cmH
2
O / mbar
(Flow)
6 to 100 L/min
t
I
(Inspiratory Time)
0.1 to 3.0 sec
ƒ
(Frequency)
1 to 99 b/min
Ptrig (Sensitivity)
–9.9 to 0 cmH2O / mbar, pressure triggering (Patient Effort
Indicator LED blinks once each time the airway pressure
reaches the Ptrig setting.)
PEEP/CPAP
0 to 30 cmH
2
O / mbar
P support (Pressure Support)
0 to 60 cmH
2
O / mbar above baseline pressure, limited to
PEEP + Psupport ≤ 60 cmH
2
O / mbar
I:E Ratio
1:99 to 3:1
Maximum Limited Airway Pressure
(Safety Valve)
100 cmH
2
O (98 mbar)
Manual Inflation
3 sec maximum
(While button is pushed, the ventilator closes the exhalation
valve and delivers an operator controlled breath to the
patient.)
Humidifier (Optional)
19ºC to 39ºC
Airway Pressure Meter
-10 to 100 cmH
2
O/-10 to 98 mbar
Alarm Silence/Reset Button & Indicator
Pressing the button silences an audible alarm violation for
60 seconds and resets a latched alarm indicator. LED lights
to indicate that Silence is active.
ALARMS Indicators
Indicators for violated alarms blink red. When the alarm is
no longer violated, the indicator latches (stays lit). Cancel a
latched indicator by pressing the Silence/Reset button.
Int. Battery Button and Indicator
Pressing the button displays the internal battery charge level
in the airway pressure meter (Paw) window. Use only when
operating on the internal battery system for accurate
reading. LED lights to indicate internal battery system
operation and alarms.
Controls/Alarms/Monitors
Range/Selection