Specifications
SYMBOLS and SPECIFICATIONS
Operator's Manual
|
2-9
(Refill Line)
Compliance at Maximum Water Level
(Full Line)
0.33 mL/cmH
2
O / mbar @ 23ºC
Intended Use
Adult and pediatric patients whose supraglottic airway is or
is not bypassed.
Warm-Up Time
30 minutes
Gas Leakage
2 mL/min at airway pressure of 80 cmH
2
O / mbar
Humidifier Output
33.8 mg/L at a continuous flow of 10 L/min @ 39ºC
Maximum Operating Airway Pressure
100 cmH
2
O / 98 mbar
Maximum Temperature at the Patient
Wye That Triggers an Alarm
41ºC
Air / Oxygen Entrainment Mixer Specifications
(optional)
Pneumatic Requirements:
Oxygen
35 to 90 psig (2.4 to 6.2 Bar) full operating range
40 to 70 psig (2.7 to 4.8 Bar) accuracy ± .08
Air
Atmospheric pressure
F
I
O
2
Control
Adjusted continuously from 0.21 to 1.00
Continuous oxygen monitoring is required for patient
safety. FLIGHT 50 does not have a built-in alarm system to
notify user of a failure or disconnect of the oxygen source.
Oxygen Blending Bag Kit Specifications
(optional)
Pneumatic Requirements:
Oxygen
Air
0-10 L/min (calibrated)
Atmospheric pressure
F
I
O
2
Control:
F
I
O
2
, indirectly adjusted from 0.21 up to 1.00 via oxygen
flow (L/min)
Continuous oxygen monitoring is required for patient
safety. FLIGHT 50 does not have a built-in alarm system
to notify user of a failure or disconnect of the oxygen
source.