Page 200
Operator‘s Manual FERITSCOPE
®
FMP30
Malfunctions and Messages
Two-point calibration
performed with calibra-
tion standards with
-
fer-
rite contents that are not
sufficiently far apart (will
be considered a one-
point calibration).
Repeat the corrective
calibration with suitable
calibration standards.
The difference of the
normalized count Xn of
the two calibration stan-
dards must be greater
than 0.1:
Xn
Calibration standard 2
- Xn
Calibration standard 1
Xn > 0,1
A probe of the same
type but with a different
serial number has been
connected.
--
Malfunction/Message Cause
Correction
start
on
Page
W
005
1-Point
calibration !
W
006
Probe changed !