Page 56
Operator‘s Manual FERITSCOPE
®
FMP30
Applications
5.7
Linking Applications
(
10.6.4 ‘Linking Applications’, beginning on Page 172)
If the linking mode is enabled, all applications that have set up using the
same probe (identified by the same serial number) will be linked with each
other. Applications that have been linked to each use the same normaliza-
tion and/or corrective calibration to determine measurements.
Example:
If a separate Application has been set up for each batch of a component
(using the same probe each time), it is useful to link the Applications to each
other in order to compare the Applications to each other. Because these are
the same kind of components, only from different batches, the same normal-
ization and/or corrective calibration should be used every time.
5.7.1
Application Linking Procedure
1.
Enable the instrument function “Linking mode” (
Applications’, beginning on Page 172).
2.
Perform a normalization to set up all Applications to be linked. The
measurement probe is assigned to the Application (
Up an Application’, beginning on Page 37).
3.
Perform a careful normalization or calibration on a typical uncoated
component (
6.2 ‘Normalization’, beginning on Page 61 and
6.3 ‘Corrective Calibration’, beginning on Page 64).
This as well as all subsequent normalizations or calibrations will be
applied o all linked Applications; the normalization stored when set-
ting up the respective Application will be overwritten.
I
5.7.2
Enabling and Disabling the Linking Mode
The linking mode is enabled or disabled from the Service functions of the
Applications that have been set up with different probes of the same
probe model (indicated by the same model designation but different
serial numbers) will not be linked.