Page 124
Operator‘s Manual FERITSCOPE
®
FMP30
Evaluation
8.2
Evaluation of the Open Application “Final Result”
The final result combines the results of the evaluation of all readings con-
tained in the open Application.
Taking a new measurement or pressing the
ENTER
key during the evalua-
tion of the open Application ends the final result display.
When making measurements with a fixed block size and the last block has
not been concluded, the next reading will be added to this block (even if
ENTER
is used to end the evaluation). In this case, the block will be con-
cluded only after the number of readings in this block corresponds to the
set block size (
7.3.6 ‘Measurements With a Fixed Block Size’, begin-
Highest value
Highest block mean value (Maximum)
LSL/USL
Upper / lower specification limit
*
Reading is within the tolerance limits
<< / >>
Reading violates the tolerance limits
!/!!
The measurement was recognized as an outlier
With the Matrix measuring mode enabled (indicated by
on the
display), the block result will
not
be printed by pressing
BLOCK-
RES
. In the Matrix measuring mode, the block result can be printed
only later by using the
command.
When making measurements with a fixed block size, only the mea-
surements of the closed blocks will be included in the evaluation of
the open Application.