Malfunctions and Messages
Instrument-internal error. Delete the faulty Appli-
cation and output the
measurement data via
the interface port.
(
[Delete Appl.?
Yes: DEL N.:EN-
TER]
:
DEL
to deleted
the Application.
[To port?
Yes: DEL N.:EN-
TER]
:
DEL
to output
the measurement data
of this Application via
the interface port to a
printer or a PC.)
If this error occurs re-
peatedly: Inform cus-
tomer service.
Unable to compute mea-
surement because after
enabling the linking
mode, no normalization
or corrective calibration
has been performed
with the connected
probe.
Perform a normaliza-
tion or corrective cali-
bration with the
connected probe.
Probe contents does not
correspond to the test
value.
Inform customer ser-
vice.
--
Malfunction/Message Cause
Correction
start
on
Page
E
025
Error in
application !
E
029
Missing
normalization
or calibration !
E
030
Probe damaged !