Page 62
Operator‘s Manual FERITSCOPE
®
FMP30
Normalization, Calibration and Master Calibration
6.2.1
Normalization Procedure
Key sequ. /
Action
Detail of the display Explanation
ZERO
Use
ZERO
to start the normalization of the
open Application.
ZERO
appears and remains on the display
while the normalization is performed.
[s]
: Standard deviation
[n]
: Number of measurements
[Base material (Fe]
: The measurements
should be made on the base material of the
calibration standard set.
[Cancel: ENTER]
: Using
ENTER
cancels
the normalization.
Base
Make about 5 measurements at different lo-
cations of the base.
The mean value of all readings obtained for
the normalization will be displayed.
[Delete: DEL]
: Use
DEL
to delete the last
measurement, 2x
DEL
to delete all read-
ings obtained for the normalization.
[OK: ENTER]
: Using
ENTER
ends the nor-
malization.
ENTER
A confirmation indicating that the normal-
ization has been carried out successfully
appears.
Pressing
ENTER
confirms the message.
: A print form of the normalization
will be printed if a printer is connected and
switched on (