Operator‘s Manual FERITSCOPE
®
FMP30
Page 117
Evaluation
8
Evaluation
The following options are available for evaluating the measured
-ferrite
contents:
Evaluation of the current block (
Block result
)
Evaluation of the open Application (
Final result
)
An evaluation cannot be carried out in an Application, where no measure-
ments have not yet been stored or where the measurements have been de-
leted! In this case, the display will not change after pressing
BLOCK-RES
or
FINAL-RES
.
The following parameters will be determined during the evaluation from the
readings of the current block, or the open Application, respectively, and
can be displayed in succession:
Block or Application number
Number of evaluated readings
Date and time of the block closure of the evaluated block or the current
date (if the evaluated block is not yet closed).
Mean value
Standard deviation or estimated standard deviation (the estimated stan-
dard deviation is shown in the final evaluation only when measure-
ments are made with a fixed block size)
Standard deviation sa (is shown in the final evaluation only if the
spread of the mean values cannot be explained from the spread within
the blocks according to a variance analysis).
Coefficient of variation V
Lowest and highest reading or lowest and highest block mean value (if
automatic block creation is enabled)
Range R
Number of measurements that are violating the lower or upper specifi-
cation limits (only if tolerance limits are enabled).
In addition, the following will be printed when outputting the block or final
results to a printer: