Operator‘s Manual FERITSCOPE
®
FMP30
Page 91
Measuring
7.3.5
Measurements With Tolerance Limits Enabled
Having the tolerance limits enabled allows for a simple and quick determi-
nation if the measured
-ferrite contents are within specified limits
(
If the measurement acquisition signal has been disabled, only the signal,
or the signals, for violating the tolerance limits will sound.
With outlier rejection enabled, the audible signals indicating an outlier
measurement will sound when an outlier is recognized. In this case, the vi-
olation of the tolerance limit will not be indicated acoustically.
7.3.6
Measurements With a Fixed Block Size
When making measurements with a fixed block size,
the block creation is carried out automatically after an adjustable num-
ber of measurements has been performed by the instrument (Setting the
Block Size:
appears on the display after the last measurement of the block has
been taken, and
As long as the tolerance limits are enabled,
will appear on the dis-
play.
Detail of the display
Explanation
After a measurement is taken that violates the lower spec-
ification limit,
appears in front of the reading on the
display. In addition, a brief audible signal sounds after
the measurement acquisition to indicated the violation of
the tolerance limit.
After a measurement is taken that violates the upper spec-
ification limit,
appears in front of the reading on the
display. In addition, two brief audible signals sound after
the measurement acquisition to indicate the violation of
the tolerance limit.