Page 100
Operator‘s Manual FERITSCOPE
®
FMP30
Measuring
If automatic block creation is enabled (
5.6.3 ‘Automatic Block Size and
Block Creation’, beginning on Page 49), the readings and the block results
of all blocks of the open Application will be printed in succession:
Fig. 7-7 Printout of the block results (example)
Application No. 0 Block No.: 1
n LSL USL
10.2% 10.8%
1| | * | | 10.4
2| | *
3| | * | | 10.3
4|<<| | | 3.1
5| | * | | 10.3
FISCHER FERITSCOPE FMP30 24.07.08
Appl.No. 0 -- Block result --
Block No.: 1 24.07.08 22:05
Mean value fe. = 8.89 +/- 4.04%
Std. dev. s = 3.25% n = 5
C. O. V. = 36.58%
lowest meas.= 3.07%
highest meas.
= 10.36%
1 values < LSL( 10.20% )= 20.00%
0 values > USL( 10.80% )= 0.00%
Application No. 0 Block No.: 2
n LSL USL
10.2% 10.8%
1| | * | | 10.3
2| | * | | 10.4
3| | * | | 10.4
4| |* | | 10.2
5| |* | | 10.2
FISCHER FERITSCOPE FMP30 24.07.08
Appl.No. 0 -- Block result --
Block No.: 2 24.07.08 22:05
Mean value fe. = 10.29 +/- 0.10%
Std. dev. s = 0.08% n = 5
C. O. V. = 0.79%
lowest meas.= 10.21%
highest meas. = 10.38%
0 values < LSL ( 10.20% )= 0.00%
0 values > USL ( 10.80% )= 0.00%
Application No. 0 Block No.: 3
n LSL USL
10.2% 10.8%
1| | * | | 10.3
2|<<| | | 10.2