Page 198
Operator‘s Manual FERITSCOPE
®
FMP30
Malfunctions and Messages
Current consumption of
a connected USB de-
vice too high (max. 100
mA permitted).
Check and/or replace
connected USB device.
--
The measurement elec-
tronics has not been bal-
anced properly.
Because of this, mea-
surement data may be
incorrect.
Inform customer ser-
vice.
--
Internal write/read er-
ror.
Inform customer ser-
vice.
--
The probe contains an
unknown data format.
Perform a software up-
date.
--
Unable to compute a
measurement because a
normalization has not
been carried out in this
Application on the cur-
rent base material.
Perform a normaliza-
tion.
Internal instrument error. Inform customer ser-
vice.
--
Malfunction/Message Cause
Correction
start
on
Page
E
031
USB overcur-
rent !
E
032
Measurem. equip.
not adjusted !
E
033
Device damaged
E
034
Unsupported
Probe format !
E
035
Please perform
a normalization
E
999
System error !!!