Operator‘s Manual FERITSCOPE
®
FMP30
Page 95
Measuring
the standards from the scope of supply of the connected probe.
Certificates in which the tolerance or the measurement uncertainty
(k=2) for the used standards is specified.
Accuracy
The calibration accuracy is limited by the measurement uncer-
tainty of the standard. The measurement uncertainty of the
corrective calibration may not be smaller than the measure-
ment uncertainty of the standards used. To improve the cali-
bration accuracy, you must use standards with smaller
measurement uncertainties for the corrective calibration.
Operating temperature of the instrument
Before a control measurement is carried out, the instrument
with connected probe must have been switched on for at least
10 minutes! This is necessary to ensure that all components in
the unit have reached their operating temperature. The result
will otherwise be mismeasurement!
For this purpose, disable automatic shutoff of the instru-
ment; see page 151.
Connect the instrument with its power supply to the
power source; see page 22.