Page 172
Operator‘s Manual FERITSCOPE
®
FMP30
Instrument Settings - Service Function Menu
10.6.4
Linking Applications
Key sequ. /
Action
Detail of the display Explanation
Select
Device Mode
by pressing the arrow
keys
or
and confirm the selection with
ENTER
.
ENTER
Select
Linking
by pressing the arrow key
or
and confirm the selection with
EN-
TER
.
ENTER
This settings option appears only when “du-
al method” is set for all Applications set up
with dual probes (
ginning on Page 37).
Select the desired mode by pressing the ar-
row keys
or
and confirm the selection
with
ENTER
.
[off]
: Applications are not linked to each
other (linking mode disabled)
[on]
: All applications that have been set up
using the same probe will be linked with
each other (linking mode enabled)
Use
DEL
at any time to cancel the setting
procedure.
As long as the linking mode f the Application is enabled
will appear on the display.
To ensure that the Applications linked to each other use the same nor-
malization and corrective calibration, a normalization or corrective
calibration, respectively, must be performed after linking for all
probes for which more than one application has been set up.