64
d3
= lb: - 0 = 1; 1 = 2; 2 = 5; 3 = 10; 4 = 20; 5 = 50
Sf
Division 2 of kg, g, lb
d1
= kg: - 0 = 1; 1 = 2; 2 = 5; 3 = 10; 4 = 20; 5 = 50
d2
= g: - 0 = 1; 1 = 2; 2 = 5; 3 = 10; 4 = 20; 5 = 50
d3
= lb: - 0 = 1; 1 = 2; 2 = 5; 3 = 10; 4 = 20; 5 = 50
Sg
Gravity Factor of
Calibration Place and
Operation Place
d1
= gravity factor of calibration place. Data length = 8 including decimal with leading zero.
d2
= gravity of location of operation place. Data length = 8 including decimal with leading zero.
Sh
Linearity
Compensation
Function
0 = Off; 1 = On
Si
ad value of zero point
(offset) value, weight
value of LD1, ad
value of LD1, weight
value of LD2 and ad
value of LD2
d1
= ad value of zero point. Data length = 8 including decimal with leading zero.
d2
= weight value of LD1. Data length = 8 including decimal with leading zero.
d3
= ad value of LD1. Data length = 8 (integers only) with leading zero.
d4
= weight value of LD2. Data length = 8 including decimal with leading zero.
d5
= ad value of LD2. Data length = 8 (integers only) with leading zero.
Sj
ad value of zero point
(offset), span weight
value, net span AD
value
d1
= as value of zero point. Data length = 8 (integers only) with leading zero.
d2
= span weight value. Data length =8 including decimal with leading zero.
d3
= net span AD value. Data length = Data length = 8 (integers only) with leading zero
Sk
Weight unit
enable/disable for kg,
g and lb
d1
= kg: - 0 = Off; 1 = On
d2
= g: - 0 = Off; 1 = On
d3
= lb: - 0 = Off; 1 = On
Sl
Filter strength
0 = level 1; 1 = level 2; … ;7 = level 8; 8 = level 9
Sm
Initial Zero range,
manual zero range,
auto zero tracking
speed
d1
= initial zero range: - 0 = Off, 1 = 1%, 2 = 2%; … 5 = 5%; 6 = 10%; 7 = 20%
d2
= manual zero range: - 0 = 1%; 1 = 2%; 2 = 3%; 3 = 4%; 4 = 5%; 5 = 10%; 6 = 20%; 7 = 50%; 8 = 75%; 9 =
100%
d3
= auto zero tracking speed: - 00 = Off; 01 = 0.25e; 02 = 0.50e; 03 = 10e; 04 = 1.5e; 05 = 2.0e; 06 = 2.5e; 07 =
3.0e; ….; 09 = 5e; 10 = 7.5e; 11 = 10e
Sn
Auto tare, repetitive
tare and preset tare
d1
= auto tare: - 0 = Off; 1 = On; 2= Continuous
d2
= repetitive tare: - 0 = Off; 1 = On
d3
= Preset tare: - 0 = Off; 1 = On
So
Stability control of
manual zero and
d1
= manual zero stability control: - 0 = No (disable); 1 = Yes (enable)
d2
= manual tare stability control: - 0 = No (disable); 1 = Yes (enable)