24
6.4.2 Auto tare (F12)
6
3 parameters are available: - Off, Auto and Contin
a.
Off: - Auto tare disable.
b.
Auto: - instrument will assume the first stable weight (≥ 20d or 20d1) applied is a container and will
then tare off the weight of it automatically. When container is removed and gross weight result is
zero, tare effect will be cancelled automatically.
c.
Contin (continuous auto tare): - all stable weight (≥ 20d or 20d1) applied will be tare off automatically.
When all loads are removed and gross weight result is zero, tare effect will be cancelled
automatically.
Notes: -
•
If Contin is set, select also d.t. (delay time, 0.0 ~ 9.9 second. Default = 0.5 second).
•
Delay time is the time duration from when a stable weight is detected until it is automatically tare
off.
•
Enter the preferred d.t. value through numeric keys and then press
[PRINT/M+]
to save.
6.4.3 Repetitive tare (F13)
7
When F13 is set to OFF, this instrument does not permit multiple tare operation. Tare effect can only be
cancelled when container is removed and gross weight is zero.
When F13 is set to ON, this instrument will permit multiple tare operation provided that
both
of the below
requirements are met: -
a.
The tare operation does not permit a reduction of the value of the tare, and
b.
The tare effect can only be cancelled when there is no load on the platform.
6.4.4 Preset tare (F63)
8
9
Preset tare allows a pre-determined tare weight value can be entered via numeric keys.
During weighing mode and when weight is zero, press
[0]
, then enter the pre-determined tare weight value
through numeric keys then press
[PRINT/M+]
to enter.
After the pre-determined tare value has been entered: -
•
Instrument displays the preset tare value entered.
•
Net indicator appears to indicate the value being displays is net weight.
To cancel preset tare effect: -
•
Remove all loads from platform then press
[TARE]
, or
•
Enter a zero preset tare value then press
[PRINT/M+]
.
Notes: -
a.
The pre-determined tare weight entered will be rounded to the nearest division of the instrument.
This does not affect the accuracy of the subsequent weighing and operation.
b.
Manual tare is possible when Repetitive Tare (F13) is set to ON.
6
Set F12 = ON to enable Auto Tare Function.
7
Set F13 = ON to enable Repeated Tare Function.
8
Set F63 = ON to enable Preset Tare Function. Some countries may not consider preset tare function as a legal for trade function.
Contact your dealer for more information.
9
Set also F12 to Off.