20
are available (refer to
Appendix B1
and
B2
for details) Press
[FUNC/SET]
or
[MR/UNIT]
until the
preferred parameter appears then press
[PRINT/M+]
to save.
5.
Instrument displays time interval (in second) between each output. 10 parameters (0, 0.5, 1.0, 1.5, 10,
30, 60, 90, 120, and 300) are available. 0 = continuous output. Press
[FUNC/SET]
or
[MR/UNIT]
until
the preferred parameter appears then press
[PRINT/M+]
to save.
6.
At this point, PC setup is completed.
5.5.2 When comport is set as CMD
1.
Instrument displays baud rate. 9 parameters (1200~256000) are available. Press
[FUNC/SET]
or
[MR/UNIT]
until the preferred parameter appears then press
[PRINT/M+]
to save.
2.
Instrument displays Parity. 3 parameters (None, odd, even) are available. Press
[FUNC/SET]
or
[MR/UNIT]
until the preferred parameter appears then press
[PRINT/M+]
to save.
3.
Instrument displays Data length. 2 parameters (7, 8) are available. Press
[FUNC/SET]
or
[MR/UNIT]
until the preferred parameter appears then press
[PRINT/M+]
to save.
4.
At this point, CMD setup is completed.
5.5.3 When comport is set as Manual
1.
Instrument displays baud rate. 9 parameters (1200~256000) are available. Press
[FUNC/SET]
or
[MR/UNIT]
until the preferred parameter appears then press
[PRINT/M+]
to save.
2.
Instrument displays Parity. 3 parameters (None, odd, even) are available. Press
[FUNC/SET]
or
[MR/UNIT]
until the preferred parameter appears then press
[PRINT/M+]
to save.
3.
Instrument displays Data length. 2 parameters (7, 8) are available. Press
[FUNC/SET]
or
[MR/UNIT]
until the preferred parameter appears then press
[PRINT/M+]
to save.
4.
Instrument displays Auto Accumulation. 2 parameters (on, off) are available. Press
[FUNC/SET]
or
[MR/UNIT]
until the preferred parameter appears then press
[PRINT/M+]
to save.
•
On = when pressing
[PRINT/M+]
during normal operation, the instrument does print and M+ at
the same time,
•
Off = when pressing
[PRINT/M+]
during normal operation, the instrument does print only.
5.
Instrument displays Check Control. 2 parameters (on, off) are available. By default, print data will only
be transmitted under all auto print modes. Press
[PRINT/M+]
to save.
•
On = (When check function is in effect) Only OK value (value which is within Lo and Hi Limits)
will be transmitted.
•
Off = (When check function is in effect) Check requirement is disable.
6.
Instrument displays Stability control. 2 parameters (Yes, no) are available. Press
[FUNC/SET]
or
[MR/UNIT]
until the preferred parameter appears then press
[PRINT/M+]
to save.
•
Yes =
[PRINT/M+]
(during normal operation) will only function when the weight is stable.
•
No =
[PRINT/M+]
(during normal operation) will always function disregarding the stable
condition of the weight when
[PRINT/M+]
is pressed.
7.
Instrument displays minimum output weight. 21 parameters (0d~20d) are available. Instrument will not
generate any output if the actual weight is less than the minimum output weight selected here. Press
[FUNC/SET]
or
[MR/UNIT]
until the preferred parameter appears then press
[PRINT/M+]
to save.
8.
Instrument displays print format. 7 parameters (Lab 1, Lab 2, Lab 3, Lab 4, Lab 5, LP-50 and TSC).
Press
[FUNC/SET]
or
[MR/UNIT]
until the preferred parameter appears then press
[PRINT/M+]
to
save.
•
Lab 1 = Output in Landscape direction. If Lab 1 is selected, refer to
5.5.3.1
for other settings.
•
Lab 2 = Output in Portrait Direction. If Lab 2 is selected, refer to
5.5.3.2
for other settings.
•
Lab 3 = Database output mode.