25
6.5 Select the Preferred Function Mode
10
This instrument supports the below function modes. Abbreviation of each function mode is bracketed.
•
Weighing (Weigh).
•
Piece Count (Count).
•
Action-Tare-Memory (AtM).
•
Animal Weighing (Ani).
•
Quick access to Customer & Product Code Setting (PCd).
Press
[FUNC/SET]
until the abbreviation of the desired function mode appears then press
[PRINT/M+]
to enter.
The working mode employed before powering off will be employed again automatically when re-powered on.
6.6 Weight Units
11
This instrument supports kg, g and lb. To shift among various weight units, press
and hold
[MR/UNIT]
to shift
among various weight units.
6.6.1 Weight unit gram (g)
Disregarding to the setting of
F9
, weight unit gram (g) is available only when 3 or 4 decimal places (0.000 or
0.0000) is selected in F81. Contact your dealer for more information about this.
6.6.2 Select the preferred weight unit
12
The desired weight units should be enabled in F9.
6.7 Memory Accumulation
There are 2 types of memory accumulation: -
a.
Automatic accumulation mode, and
b.
Manual Accumulation mode.
Maximum accumulation limit is = 8 digits (e.g. 99999999) plus decimal (if any).
Err 28
appears when
maximum accumulation limit is exceeded.
6.7.1 Automatic Accumulation
13
Automatically accumulation is activated when Auto Accumulation is set to
On
in Auto 1, Auto 2, Auto 3 or Manual
mode is selected in F16 and/or F17.
Under the automatic accumulation mode, corresponding results will be accumulated automatically.
6.7.2 Manual Accumulation
14
Manual Accumulation is activated when scanner, PC or CMD mode is selected for
both
F16 and/or F17.
Under the Manual Accumulation mode, press
[PRINT/M+]
to accumulate the current value to memory.
10
Depends on F11 setting.
11
Depends on F9 setting.
12
Changing weight unit during operation will clear all accumulate weight data from memory.
13
Refer to
5.5.3 & 5.5.4
for setting details.
14
Refer to
5.5.3 & 5.5.4
for setting details.